NIMBUS
Report
- Report Number
- 3005619970-2018-00020
- Event Type
- Death
- Date Received
- December 14, 2018
- Date of Event
- November 1, 2018
- Report Date
- January 25, 2019
- Manufacturer
- GETINGE (SUZHOU) CO., LTD.
- Product Code
- FNM
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING FILED UNDER EXEMPTION E2012066 BY THE GETINGE SUZHOU LTD (REGISTRATION #(B)(4) ON BEHALF OF THE IMPORTER ARJOHUNTLEIGH INC (AHUS) (REGISTRATION #(B)(4). ADDITIONAL INFORMATION WILL BE PROVIDED UPON CONCLUSION OF THE INVESTIGATION.
ARJO WAS NOTIFIED ABOUT AN INCIDENT INVOLVING NIMBUS SYSTEM. THE EXACT MODEL NUMBER AND SERIAL NUMBER WAS NOT PROVIDED. IT WAS REPORTED BY (B)(6) FIRE AND RESCUE SERVICE THAT THERE WAS A FATAL FIRE AND THE DECEASED WAS SLEEPING ON WHAT WAS BELIEVED TO BE A NIMBUS AIR MATTRESS. CURRENTLY, THERE IS NO INDICATION THAT ARJO MATTRESS COULD CONTRIBUTE TO THE ISSUE. INVESTIGATION BY FIRE AND RESCUE SERVICE IS ONGOING.
ARJO WAS NOTIFIED ABOUT AN INCIDENT INVOLVING NIMBUS SYSTEM. THE EXACT MODEL NUMBER AND SERIAL NUMBER WAS NOT PROVIDED. IT WAS REPORTED BY (B)(6) THAT THERE WAS A FATAL FIRE AND THE PATIENT WAS SLEEPING ON WHAT WAS BELIEVED TO BE A NIMBUS AIR MATTRESS. AS THE FIRE SERVICE DURING INSPECTION NOTICE THAT BRAND NAME "NIMBUS" WAS VISIBLE ON REMAINS OF PUMP. THE EQUIPMENT HAS BEEN ISOLATED. ARJO HAD NO ACCESS TO CONFIRMED ANY DETAILS. CURRENTLY, THERE WAS NO INDICATION THAT ARJO MATTRESS COULD CONTRIBUTE TO THE ISSUE. INVESTIGATION BY FIRE AND RESCUE SERVICE IS ONGOING. NO ADDITIONAL INFORMATION WAS PROVIDED TO THIS DATE. NIMBUS MATTRESSES COMPRISE OF AIR-INFLATED, ALTERNATING PRESSURE CELLS DEDICATED FOR PREVENTION AND MANAGEMENT OF PRESSURE ULCERS. TOP COVERS OF NIMBUS MATTRESSES ARE MADE OF A PU COATED FABRIC WHICH IS FIRE RETARDANT MATERIAL TESTED IN ACCORDANCE WITH APPLICABLE STANDARDS. PLEASE NOTE THAT MATTRESSES BASE COVERS HAS BEEN TESTED ACCORDING TO BS7175 - (B)(4) STANDARD - FLAME RETARDANT BEDDING FOR BEDCOVERS AND SHEETS AND PASSED THIS TEST WITH POSITIVE RESULT. FOLLOWING THE INSTRUCTION FOR USE #151996: "DO NOT EXPOSE THE SYSTEM, ESPECIALLY THE MATTRESS, TO NAKED FLAMES, SUCH AS CIGARETTES, ETC". "IN THE EVENT OF A FIRE, A LEAK IN THE SEAT OR MATTRESS COULD PROPAGATE THE FIRE." FROM ARJO REVIEW IT WAS CONCLUDED THAT THERE THE PRODUCT IS NOT VIOLATIVE IN ANY WAY. THE MATERIAL OF NIMBUS TOP COVER IS CONFORMING TO APPLICABLE STANDARDS AND IS NOT CONSIDERED TO HAVE CONTRIBUTED TO THE OUTCOME OF THIS UNFORTUNATE INCIDENT WHICH WAS REPORTED. TO SUM UP THERE WAS LIMITED INFORMATION RELATED TO EVENT CIRCUMSTANCES. THE NIMBUS SYSTEM PLAYED A ROLE IN THE INCIDENT AS IT IS BELIEVED THAT THE PATIENT WAS SLEEPING ON THE MATTRESS. IF THE INVESTIGATION RESULTS (CONDUCTED BY FIRE AND RESCUE SERVICE) WILL BE SHARED AND GIVE NEW LIGHT INTO THE INVESTIGATION OR ROOT CAUSE, THE FOLLOW UP REPORT WITH NEW CONCLUSION WILL BE SENT TO AUTHORITIES. COMPLAINT DECIDED TO BE REPORTABLE DUE TO A CRITICAL PATIENT'S OUTCOME.
COLLECTING INFORMATION ONGOING. ADDITIONAL INFORMATION WILL BE PROVIDED UPON CONCLUSION OF THE INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1003417 | NIMBUS | MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE | FNM | GETINGE (SUZHOU) CO., LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death |