FDA Adverse Event Injury Summary report: N

8010478-2018-00005

MDR report key: 8161565 · Received December 13, 2018

Report

Report Number
8010478-2018-00005
Event Type
Injury
Date Received
December 13, 2018
Date of Event
November 14, 2018
Report Date
November 14, 2018
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION REVEALED THE FOLLOWING: THE INCIDENT WAS USER RELATED DUE TO A WRONG APPLICATION FROM THE CUSTOMER SITE. HE DID USE A PATHOLOGICAL TISSUE QUICK FREEZING SPRAY, NAMED WHITE FREEZER S, WHICH IS A FLAMMABLE SUBSTANCE, AND NOT RECOMMENDED FOR USE IN THE LEICA CM1950, ACCORDING TO THE INSTRUCTIONS FOR USE. A CUSTOMER FACING LETTER WILL BE SENT TO THE CUSTOMER WITH A RECOMMENDATION IN THE FUTURE NOT TO USE FLAMMABLE SUBSTANCES IN THE LEICA CM1950 WHEN IT IS TURNED ON AND PLUGGED IN.

Description of Event or Problem · 1

A CUSTOMER REPORTED TO LEICA BIOSYSTEMS THAT ON (B)(6) 2018 COMBUSTION WITH FIRE OCCURRED DURING THE USAGE OF AN INSTANT COOLING AGENT ON THE FREEZING SHELF IN THEIR CM1950. ON (B)(6) 2018 THE CUSTOMER INFORMED LEICA BIOSYSTEMS THAT HE WAS BURNED ON HIS ARM AND THE HAIR ON HIS FOREHEAD. HE WAS WEARING A MASK AND GOGGLES WHICH PREVENTED HIM FROM BEING BURNED ON HIS FACE. MEDICAL TREATMENT WAS NECESSARY.

Patients

Seq Age Sex Outcome Treatment
1 Other