8010478-2018-00005
Report
- Report Number
- 8010478-2018-00005
- Event Type
- Injury
- Date Received
- December 13, 2018
- Date of Event
- November 14, 2018
- Report Date
- November 14, 2018
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION REVEALED THE FOLLOWING: THE INCIDENT WAS USER RELATED DUE TO A WRONG APPLICATION FROM THE CUSTOMER SITE. HE DID USE A PATHOLOGICAL TISSUE QUICK FREEZING SPRAY, NAMED WHITE FREEZER S, WHICH IS A FLAMMABLE SUBSTANCE, AND NOT RECOMMENDED FOR USE IN THE LEICA CM1950, ACCORDING TO THE INSTRUCTIONS FOR USE. A CUSTOMER FACING LETTER WILL BE SENT TO THE CUSTOMER WITH A RECOMMENDATION IN THE FUTURE NOT TO USE FLAMMABLE SUBSTANCES IN THE LEICA CM1950 WHEN IT IS TURNED ON AND PLUGGED IN.
A CUSTOMER REPORTED TO LEICA BIOSYSTEMS THAT ON (B)(6) 2018 COMBUSTION WITH FIRE OCCURRED DURING THE USAGE OF AN INSTANT COOLING AGENT ON THE FREEZING SHELF IN THEIR CM1950. ON (B)(6) 2018 THE CUSTOMER INFORMED LEICA BIOSYSTEMS THAT HE WAS BURNED ON HIS ARM AND THE HAIR ON HIS FOREHEAD. HE WAS WEARING A MASK AND GOGGLES WHICH PREVENTED HIM FROM BEING BURNED ON HIS FACE. MEDICAL TREATMENT WAS NECESSARY.
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |