LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2018-10561
- Event Type
- Injury
- Date Received
- December 13, 2018
- Date of Event
- November 5, 2018
- Report Date
- November 30, 2018
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- UDI-DI
- 00855778005005
- PMA / PMN Number
- P010030
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
EXPLANATION: THERE WAS NO DEATH OR DEVICE MALFUNCTION ASSOCIATED WITH THE INAPPROPRIATE DEFIBRILLATION EVENT. THER E IS NO INFORMATION TO SUGGEST THAT THE PATIENT SUSTAINED A SERIOUS INJURY. DEVICE EVALUATION SUMMARY: THE MONITOR SN (B)(4) AND ELECTRODE BELT SN (B)(4) WERE RETURNED TO THE DISTRIBUTOR FOR EVALUATION, IN ACCORDANCE WITH PROCEDURES RECOMMENDED BY ZOLL MANUFACTURING CORPORATION. THE EVALUATION INCLUDED REVIEW OF DOWNLOADED SOFTWARE FLAG FILES ON THE DAY OF THE EVENT AND INCOMING FUNCTIONAL TESTING. THE REVIEW OF THE SOFTWARE FLAGS CONSISTED OF AN ANALYSIS OF THE DOWNLOADED DATA TO IDENTIFY ANY FAULT FLAGS OR UNUSUAL PATTERNS OF SOFTWARE FLAGS. THE SOFTWARE FLAG FILES DID NOT SUGGEST A DEVICE MALFUNCTION THAT WOULD CONTRIBUTE TO THE INAPPROPRIATE TREATMENT. DURING THE INCOMING FUNCTIONAL TESTING, A 1HZ SIMULATED NORMAL SINUS RHYTHM SIGNAL WAS APPLIED TO THE ECG ELECTRODES, FOLLOWED BY A 5HZ SIMULATED TREATABLE ARRHYTHMIA SIGNAL WHICH VERIFIED PROPER PERFORMANCE OF THE DETECTION ALGORITHM. DURING THE TRANSITION TO THE 5HZ SIGNAL, THE DEVICE WAS CONFIRMED TO PROPERLY ENTER INTO A TREATMENT SEQUENCE WHICH INCLUDES A VERIFICATION OF THE TACTILE VIBRATION ALARM, AUDIO MESSAGING, AND SIREN ALARMS, AS WELL AS A TEST OF THE PULSE DELIVERY CIRCUITRY. THE PULSE DELIVERY CIRCUITRY TEST VERIFIED PROPER CHARGING OF THE HIGH VOLTAGE CAPACITORS AND PROPER DELIVERY OF FIVE FULL ENERGY 150J BIPHASIC PULSES. THE FUNCTIONAL TESTING CONFIRMED PROPER RESPONSE BUTTON FUNCTIONALITY, ECG ACQUISITION, DETECTION ALGORITHM PERFORMANCE, AND PULSE DELIVERY FUNCTIONALITY. THE INVESTIGATION INTO THE EVENT CONCLUDES THAT THERE WAS NO DEVICE MALFUNCTION. A CAUSE AND EFFECT ANALYSIS WAS CONDUCTED USING ALL OF THE AVAILABLE INFORMATION WHICH INCLUDES THE INCIDENT REPORT, DEVICE EVALUATION, SOFTWARE FLAG FILES, AND ECG STRIPS. THE PRIMARY CAUSE OF THE INAPPROPRIATE SHOCK EVENT WAS LACK OF RESPONSE BUTTON USE PRIOR TO SHOCK DELIVERY. THE ECG ANALYSIS, CONDUCTED BY TRAINED ECG TECHNICIANS, IDENTIFIED THE PRIMARY CAUSE OF THE FALSE DETECTION WAS MOTION ARTIFACT. THE SOURCE OF THE MOTION ARTIFACT COULD NOT BE POSITIVELY IDENTIFIED THROUGH THE CAUSE AND EFFECT ANALYSIS. THE FOLLOWING FACTORS COULD NOT BE RULED OUT AS CONTRIBUTING CAUSES OF THE MOTION ARTIFACT: BODY MOTION, POOR ECG CONTACT WITH SKIN. INAPPROPRIATE DEFIBRILLATIONS ARE AN ANTICIPATED RISK ASSOCIATED WITH THE USE OF THE LIFEVEST. PATIENTS ARE INSTRUCTED THROUGH ALARMS, VOICE MESSAGES, IFU, AND TRAINING TO PRESS THE RESPONSE BUTTONS TO PREVENT AN INAPPROPRIATE DEFIBRILLATION. (B)(4). A SUMMARY OF THE SAFETY AND EFFECTIVENESS DATA (SSED), INCLUDING THE INAPPROPRIATE DEFIBRILLATION SAFETY OBJECTIVE SUPPORTING FDA'S APPROVAL OF THE LIFEVEST, CAN BE FOUND AT HTTP://WWW.ACCESSDATA.FDA.GOV/CDRH_DOCS/PDF/P010030B.PDF. THE LIFEVEST DETECTION ALGORITHM COMPLIES WITH IEC 60601-2-4 PERFORMANCE REQUIREMENTS FOR SENSITIVITY AND SPECIFICITY.
A US DISTRIBUTOR CONTACTED ZOLL TO REPORT THAT A PATIENT EXPERIENCED AN INAPPROPRIATE TREATMENT WHILE IN A SKILLED NURSING FACILITY. MOTION ARTIFACT CONTRIBUTED TO THE FALSE DETECTION. THE PATIENT WAS CONSCIOUS AT THE TIME OF THE EVENT BUT DID NOT PRESS THE RESPONSE BUTTONS PRIOR TO THE TREATMENT. THE PATIENT DID NOT SEEK MEDICAL ATTENTION AND CONTINUED TO WEAR THE LIFEVEST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1001988 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 | 00855778005005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Other |