ALERE DETERMINE HIV 1/2 AG/AB COMBO
Report
- Report Number
- 1221359-2018-00551
- Event Type
- Malfunction
- Date Received
- December 13, 2018
- Date of Event
- November 20, 2018
- Report Date
- December 13, 2018
- Manufacturer
- ALERE SCARBOROUGH INC.
- Product Code
- MZF
- UDI-DI
- 10811877010293
- PMA / PMN Number
- BP120037-0
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
TESTING WAS PERFORMED AT ALERE (B)(4) ON RETAINED KIT LOTS 102459 AND 101439 WITH THE FOLLOWING INTERNAL SERUM PLASMA CONTROL SAMPLES: (B)(6). ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING BATCH RECORDS FOR LOTS 102459 AND 101439 WERE REVIEWED. THESE LOTS MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS (B)(6) RELATED TO LOTS 102459 AND 101439 SHOWED THAT THE COMPLAINT RATE IS (B)(4), RESPECTIVELY. THE EVIDENCE AVAILABLE DOES NOT INDICATE THAT THE PRODUCT IS PERFORMING OUTSIDE LABEL CLAIMS. ALERE (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. THE RESULTS OBTAINED MAY POSSIBLY BE RELATED TO THE PATIENT SAMPLE. THE SAMPLE MAY HAVE CONTAINED SPECIFIC SUBSTANCES WHICH MAY HAVE AFFECTED THE RESULTS. THE AVAILABLE EVIDENCE SUGGESTS THAT THESE DEVICE LOTS ARE PERFORMING WITHIN LABELED CLAIMS.
A CUSTOMER REPORTED (B)(6) RESULTS ON A SERUM PLASMA SAMPLE WITH THE ALERE DETERMINE HIV 1/2 AG/AB COMBO. A TOTAL OF 4 TEST STRIPS FROM 2 DIFFERENT LOTS WERE TESTED FOR THE SINGLE REPORTED PATIENT. THERE IS INSUFFICIENT INFORMATION TO DETERMINE IF A MALFUNCTION OCCURRED. THE RESULTS OF A (B)(6) TEST WERE UNKNOWN AT THE TIME OF REPORTING. IT WAS INDICATED THAT THE FEMALE PATIENT WAS PREGNANT. THE PATIENT'S TREATMENT AND OUTCOME WERE UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 999582 | ALERE DETERMINE HIV 1/2 AG/AB COMBO | IVD FOR HIV AG/AB | MZF | ALERE SCARBOROUGH INC. | 102459 | 10811877010293 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR |