VITEK® 2 AST-P626 TEST KIT
Report
- Report Number
- 1950204-2018-00641
- Event Type
- Malfunction
- Date Received
- December 12, 2018
- Report Date
- April 25, 2019
- Manufacturer
- BIOMERIEUX, INC.
- Product Code
- LON
- PMA / PMN Number
- K053097
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
AN INTERNAL BIOMÉRIEUX INVESTIGATION WAS PERFORMED FOR A CUSTOMER IN (B)(6) DUE TO A REPORT OF FALSE POSITIVE CEFOXITIN SCREEN (OXSF) RESULTS FOR TWO PATIENT STRAINS OF STAPHYLOCOCCUS AUREUS ASSOCIATED WITH VITEK® 2 SYSTEM V8.01 SOFTWARE, AND VITEK® 2 AST-P626 CARD. THE LAB REPORTS RECEIVED FROM THE CUSTOMER FOR TESTING PERFORMED USING THE VITEK 2 AST-P626 CARDS WERE FROM LOTS: 5360458103 (EXPIRED), 5360424103 (EXPIRED), AND 5360438103 (EXPIRED). SINCE CUSTOMER LOTS (CL: 5360723103 AND CL: 5360662103) WERE NOT EXPIRED AT THE TIME OF THIS INVESTIGATION AND HAD EACH BEEN USED BY THE CUSTOMER, THESE LOTS WERE SELECTED ALONG WITH A RANDOM LOT (RL: 5360837203), FOR TESTING OF THIS INVESTIGATION. THE CUSTOMER'S STRAINS WERE AS FOLLOWS: 67168952 (911903) AND 62153680 (911904). IDENTIFICATION WAS CONFIRMED FOR BOTH OF THE CUSTOMER'S STRAINS AS STAPHYLOCOCCUS AUREUS VIA VITEK® MS, MATRIX ASSISTED LASER DESORPTION IONIZATION-TIME OF FLIGHT (MALDI-TOF) MASS SPECTROMETRY. THE FOLLOWING REFERENCE METHODS WERE USED TO DETERMINE THE INTENDED RESULT: · KIRBY-BAUER CEFOXITIN (FOX KB), REFERENCE METHOD USED FOR CEFOXITIN SCREEN TEST (OXFS01N) WITH CLSI DIAMETERS BREAKPOINTS USED IN DEVELOPMENT [>= 22 MM S; <= 21 MM R] · OXACILLIN (OX) AGAR DILUTION (AD), REFERENCE METHOD USED FOR THE DEVELOPMENT OF OX ON THE AST-P626 CARDS ADDITIONAL TESTING VIA THE PBP2A METHOD WAS CONDUCTED TO DETERMINE THE PRESENCE OF (B)(6). THE RESULTS FOR REFERENCE TEST METHODS FOX KB AND OX AD, AND THE ADDITIONAL TEST METHOD PBP2A WERE AS FOLLOWS: 67168952 (911903): FOX KB DIAMETER = 26 MM (S), AD OX MIC = 0.25 MG/L, PBP2A NEGATIVE. 62153680 (911904): FOX KB DIAMETER = 24 MM (S), AD OX MIC = 1 MG/L, PBP2A NEGATIVE. ANTIBIOTIC SUSCEPTIBILITY TESTING WAS PERFORMED USING THE VITEK 2 SYSTEM SOFTWARE VERSION V08.01, AND THE VITEK 2 AST-P26 CARD. THREE AST-P626 CARD LOTS (MENTIONED ABOVE), INCLUDING TWO CUSTOMER LOTS (CL: 5360723103 AND CL: 5360662103) AND A RANDOM LOT (RL: 5360837203) WERE TESTED. REMEL TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD WAS USED TO PERFORM SUBCULTURE. THE OXSF RESULTS FOR BOTH THE CUSTOMER LOTS, AND THE RANDOM LOT RESULTED IN NEGATIVE CEFOXITIN SCREEN TESTS FOR ALL CARDS TESTED FOR BOTH STRAINS TESTED; HOWEVER, ISOLATE 62153680 (911904), DID RESULT IN A RESISTANT OXACILLIN MIC >= 4.0 MG/L, WHICH LED TO A BIOLOGICAL CORRECTION TO THE CEFOXITIN SCREEN FORCING IT FROM NEGATIVE TO POSITIVE. THE OX AND OXFS RESULTS FOR ISOLATE 67168952 (911903) WERE IN BOTH ESSENTIAL AND CATEGORY AGREEMENT. THE VITEK 2 RESULTS FOR OXFS AND OX ARE LISTED BELOW: 67168952 (911903), OXSF NEGATIVE (3 RESULTS), OX MIC = 0.5 MG/L (1 RESULT) AND MIC <= 0.25 MG/L (2 RESULTS) 62153680 (911904), *OXSF NEGATIVE BUT CORRECTED TO POSITIVE (3 RESULTS), OX MIC >= 4.0 MG/L (3 RESULTS) IN CONCLUSION, THE CUSTOMER'S FALSE POSITIVE CEFOXITIN SCREEN RESULTS WERE NOT REPRODUCED INTERNALLY. ISOLATE 62153680 (911904), YIELDED A RESULT OF OX MIC >= 4.0 MG/L AND LEAD TO A BIOLOGICAL CORRECTION OF THE CEFOXITIN SCREEN.
A CUSTOMER IN (B)(6) REPORTED A FALSE POSITIVE CEFOXITIN SCREEN FOR STAPHYLOCOCCUS AUREUS STRAINS A IN ASSOCIATION WITH THE VITEK® 2 AST-P626 TEST KIT. THE CUSTOMER FIRST REPORTED THE ISSUE WHEN TESTING THE STAPHYLOCOCCUS AUREUS (ATCC 29213) INTERNAL QUALITY CONTROL STRAIN, BUT NOW REPORTS THE CEFOXITIN SCREEN WAS POSITIVE AND (B)(6) FOR TWO PATIENT STRAINS ((B)(6) AND (B)(6)). THERE IS NO INDICATION OR REPORT FROM THE HOSPITAL OR TREATING PHYSICIAN TO BIOMÉRIEUX THAT THE DISCREPANT RESULT LED TO ANY ADVERSE EVENT RELATED TO THE PATIENT'S STATE OF HEALTH. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 994842 | VITEK® 2 AST-P626 TEST KIT | VITEK® 2 AST-P626 TEST KIT. | LON | BIOMERIEUX, INC. | 5360458103 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |