FDA Adverse Event Malfunction Summary report: N

VITALITY AVT

MDR report key: 815397 · Received February 8, 2007

Report

Report Number
2124215-2007-19708
Event Type
Malfunction
Date Received
February 8, 2007
Date of Event
November 16, 2006
Report Date
November 16, 2006
Manufacturer
CARDIAC PACEMAKERS, INC
Product Code
LWS
Removal / Correction Number
Z1107-9/05
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
LAY USER/PATIENT

Narratives

Description of Event or Problem · 1

EVENT DESCRIPTION BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DISPLAYED AN ELECTIVE REPLACEMENT INDICATOR (ERI). THERE IS A CONCERN THAT THE BATTERY DEPLETED EARLIER THAN EXPECTED. THE MONITORING VOLTAGE WAS 2.69V WITH A CHARGE TIME OF 26.9 SECONDS. THIS DEVICE IS PART OF THE AVT LATCHING POPULATION (6/17/2005).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY AVT IMPLANTABLE ATRIAL GENERATOR LWS CARDIAC PACEMAKERS, INC A135 NA

Patients

Seq Age Sex Outcome Treatment
1 59 YR Other