FDA Adverse Event
Malfunction
Summary report: N
R SERIES DEFIBRILLATOR
MDR report key: 8151609
·
Received December 11, 2018
Report
- Report Number
- 1220908-2018-03606
- Event Type
- Malfunction
- Date Received
- December 11, 2018
- Report Date
- November 28, 2018
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- DRT
- UDI-DI
- 00847946017200
- PMA / PMN Number
- K060559/P160
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ADDITIONAL PRO CODE: MKJ. THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY ECG SHIELD ON THE ANALOG BOARD. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED "ECG DISABLED" AND "DEFIB PAD SHORT" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 990333 | R SERIES DEFIBRILLATOR | DEFIBRILLATOR/PACEMAKER | DRT | ZOLL MEDICAL CORPORATION | R SERIES | NA | 00847946017200 |