FDA Adverse Event Malfunction Summary report: N

R SERIES DEFIBRILLATOR

MDR report key: 8151609 · Received December 11, 2018

Report

Report Number
1220908-2018-03606
Event Type
Malfunction
Date Received
December 11, 2018
Report Date
November 28, 2018
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
DRT
UDI-DI
00847946017200
PMA / PMN Number
K060559/P160
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL PRO CODE: MKJ. THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY ECG SHIELD ON THE ANALOG BOARD. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED "ECG DISABLED" AND "DEFIB PAD SHORT" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
990333 R SERIES DEFIBRILLATOR DEFIBRILLATOR/PACEMAKER DRT ZOLL MEDICAL CORPORATION R SERIES NA 00847946017200

Patients

Seq Age Sex Outcome Treatment
1