FDA Adverse Event Malfunction Summary report: N

BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER

MDR report key: 8151355 · Received December 11, 2018

Report

Report Number
1710034-2018-00904
Event Type
Malfunction
Date Received
December 11, 2018
Date of Event
November 21, 2018
Report Date
January 10, 2019
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
30382903814122
PMA / PMN Number
K952861
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: DHR REVIEW WAS PERFORMED ON LOT NUMBER 8232675; THE LOT NUMBER WAS BUILT ON AFA LINE 4 FROM 25AUG2018 THRU 30AUG2018. PACKAGED ON PACKAGING LINE 11 FROM 26AUG2018 THRU 30AUG2018 (PALLETS 1-7). PACKAGED ON PACKAGING LINE 9 FROM 30AUG2018 THRU 05SEPT2018 (PALLETS 8-11). REVIEW OF DHR REVEALED ALL REQUIRED CHALLENGE SAMPLES, SET-UP AND IN PROCESS TESTING WAS PERFORMED IN ACCORDANCE WITH THE QUALITY PLANS. OBSERVATIONS AND TESTING COULD NOT BE PERFORMED BECAUSE UNITS WERE NOT RECEIVED FOR INVESTIGATION OF THIS INCIDENT. INDETERMINATE ¿ WITHOUT THE ACTUAL SAMPLE FOR EVALUATION AND TESTING THERE WAS NO PHYSICAL EVIDENCE TO CONFIRM OR SUPPORT MANUFACTURING PROCESS RELATED ISSUES FOR THE REPORTED DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE OF THE BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER THE TIP OF THE CATHETER STARTED TO SPLIT.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE OF THE BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER THE TIP OF THE CATHETER STARTED TO SPLIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
989924 BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 8232675 30382903814122

Patients

Seq Age Sex Outcome Treatment
1 Other