FDA Adverse Event Malfunction Summary report: N

SYNCARDIA FREEDOM DRIVER

MDR report key: 8148378 · Received December 10, 2018

Report

Report Number
3003761017-2018-00546
Event Type
Malfunction
Date Received
December 10, 2018
Date of Event
November 22, 2018
Report Date
February 27, 2019
Manufacturer
SYNCARDIA SYSTEMS, LLC
Product Code
LOZ
UDI-DI
00858000003121
PMA / PMN Number
P030011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

VISUAL INSPECTION OF THE DRIVER REVEALED NO ABNORMALITIES. THE DRIVER PASSED INCOMING FUNCTIONAL TEST REQUIREMENTS. ADDITIONALLY, THE DRIVER WAS SUBJECTED TO AN EXTENDED 48 HOUR OBSERVATION RUN WHERE IT PERFORMED AS INTENDED WITH NO ABNORMALITIES OR ALARMS. THE RETURNED ONBOARD BATTERIES WERE ALSO TESTED AND ALL SUCCESSFULLY PASSED FINAL TEST ACCEPTANCE CRITERIA WITH NO ABNORMALITIES. A TEMPERATURE ALARM CAN BE PRODUCED IF THE TEMPERATURE OF THE ONBOARD BATTERIES IS TOO HOT OR TOO COLD. THE FREEDOM DRIVER SYSTEM OPERATOR MANUAL (F-900013) HAS MULTIPLE PRECAUTIONS AND WARNINGS ABOUT THE FREEDOM DRIVER'S USE IN COLD TEMPERATURES, DETAILED BELOW. - SECTION 5.1 PRECAUTIONS: "AVOID EXPOSING THE FREEDOM DRIVER SYSTEM TO EXTREME TEMPERATURES FOR MORE THAN BRIEF PERIODS OF TIME" - SECTION 10.2 TEMPERATURE ALARM: "THE ISSUE THAT CAUSED THE ALARM MUST BE IMMEDIATELY ADDRESSED BY PERFORMING THE FOLLOWING STEPS", WHICH INCLUDE MOVING THE FREEDOM DRIVER INTO A CLIMATE CONTROLLED ENVIRONMENT (A WARMER AREA). - SECTION 10.2 CAUTION: "DO NOT IGNORE A TEMPERATURE ALARM. AN UNRESOLVED TEMPERATURE ALARM MAY BECOME A PERMANENT FAULT ALARM." - SECTION 17.2 MAINTENANCE AND CLEANING INSTRUCTIONS: "DO NOT EXPOSE THE FREEDOM DRIVER TO TEMPERATURES HIGHER THAN 40°C (104°F) OR LOWER THAN 4.5°C (40°F) FOR LONG PERIODS." - SECTION 17.2.4 ONBOARD BATTERIES: "STORE THE ONBOARD BATTERIES IN A COOL, DRY PLACE AND WITHIN A TEMPERATURE RANGE OF 0°C TO 40°C (32°F TO 104°F). CHARGE ONBOARD BATTERIES WITHIN A TEMPERATURE RANGE OF 0°F TO 40°C (32°F TO 104°F)." AND "DO NOT LEAVE THE ONBOARD BATTERIES EXPOSED TO EXTREME HEAT OR COLD. EXTREME HIGH AND LOW TEMPERATURES CAN DAMAGE THE ONBOARD BATTERIES." - SECTION 18.6 ENVIRONMENTAL CONDITIONS: "OPERATING TEMPERATURE (DRIVER): 4.5° TO 40°C (40° TO 104°F). OPERATING TEMPERATURE (ACCESSORIES): 4.5° TO 40°C (40° TO 104°F). TRANSPORTATION/STORAGE TEMPERATURE: -25° TO 70°C (-13° TO 158°F)." THE REPORTED TEMPERATURE THAT THE DRIVER AND ONBOARD BATTERIES WERE EXPOSED TO WAS OUTSIDE OF THE RECOMMENDED OPERATING RANGE (BELOW 40°F FOR THE DRIVER AND 32°F FOR THE ONBOARD BATTERIES), BUT WITHIN THE TRANSPORTATION/STORAGE TEMPERATURE RANGE WHICH GOES DOWN TO -13°F. ALTHOUGH THE DRIVER AND ACCESSORIES SHOULD NOT BE OPERATING IN THESE CONDITIONS, THIS INVESTIGATION DETERMINED THAT THERE WAS NO DEVICE MALFUNCTION CAUSED BY THE DRIVER AND ONBOARD BATTERIES BEING EXPOSED TO THESE CONDITIONS. THIS ISSUE WILL CONTINUE TO BE MONITORED AND TRENDED AS PART OF THE CUSTOMER EXPERIENCE PROCESS. SYNCARDIA HAS COMPLETED ITS EVALUATION AND IS CLOSING THIS FILE. CE 4582 FOLLOW-UP REPORT 1.

Additional Manufacturer Narrative · 1

THIS ALLEGED FAILURE MODE POSES A LOW RISK TO THE PATIENT BECAUSE THE ISSUE DID NOT PREVENT THE DRIVER FROM PERFORMING ITS LIFE-SUSTAINING FUNCTION. THE FREEDOM DRIVER WILL BE RETURNED TO SYNCARDIA FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WILL BE PROVIDED IN A FOLLOW-UP MDR. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT THE PATIENT WAS AT A FOOTBALL GAME IN APPROXIMATELY 19 DEGREE FAHRENHEIT TEMPERATURES FOR 2.5 HOURS WHEN HIS FREEDOM DRIVER EXHIBITED A FAULT ALARM. PATIENT'S WIFE CHECKED THE DRIVELINES, CHANGED THE BATTERIES, WHICH WERE ALSO OUTSIDE IN THE LOW TEMPERATURES, BUT THE FAULT ALARM CONTINUED. THE CUSTOMER ALSO REPORTED THAT THE PATIENT WAS SUBSEQUENTLY SWITCHED TO THE BACKUP FREEDOM DRIVER WHICH HAD BEEN IN THE CAR. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
987542 SYNCARDIA FREEDOM DRIVER EXTERNAL PNUEMATIC DRIVER LOZ SYNCARDIA SYSTEMS, LLC 595000-001 00858000003121

Patients

Seq Age Sex Outcome Treatment
1 56 YR