FDA Adverse Event Injury Summary report: N

CONTINUUM LINER

MDR report key: 8147498 · Received December 10, 2018

Report

Report Number
0001822565-2018-06812
Event Type
Injury
Date Received
December 10, 2018
Date of Event
November 5, 2018
Report Date
September 11, 2019
Manufacturer
ZIMMER BIOMET, INC.
Product Code
JDI
PMA / PMN Number
K151448
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. REPORTED EVENT WAS UNABLE TO BE CONFIRMED DUE TO LIMITED INFORMATION RECEIVED FROM THE CUSTOMER. DEVICE HISTORY RECORD (DHR) REVIEW WAS UNABLE TO BE PERFORMED AS THE LOT NUMBER OF THE DEVICE INVOLVED IN THE EVENT IS UNKNOWN. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. REPORTED EVENT WAS CONFIRMED BY REVIEW OF IMAGES PROVIDED SHOWING DAMAGE ON THE LOCKING MECHANISM OF THE LINER AS WELL AS ON THE OUTER WALL OF THE LINER. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ADDITIONAL INFORMATION DOES NOT AFFECT THE ROOT CAUSE. ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. UDI# (01) 00889024149158 (17) 210731 (10) 63412126. THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). REPORT SOURCE (B)(6). CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION APPROXIMATELY 1.5 YEARS POST IMPLANTATION DUE TO REPEATED DISLOCATIONS. THE SURGEON OPTED TO REMOVE AND REPLACE THE HEAD AND LINER COMPONENTS. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
989025 CONTINUUM LINER PROSTHESIS, HIP JDI ZIMMER BIOMET, INC. N/A 63412126

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R UNK HEAD| UNK HEAD