DIMENSION EXL WITH LM
Report
- Report Number
- 2517506-2018-00680
- Event Type
- Malfunction
- Date Received
- December 10, 2018
- Date of Event
- November 3, 2018
- Report Date
- January 4, 2019
- Manufacturer
- SIEMENS HEATLHCARE DIAGNOSTICS INC.
- Product Code
- JJE
- PMA / PMN Number
- K130276
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SIEMENS FILED THE INITIAL MDR# 2517506-2018-00680 ON 10-DEC-2018. ADDITIONAL INFORMATION (14-DEC-2018): UDI # WAS UPDATED WITH THE UDI NUMBER. ADDITIONAL INFORMATION (02-JAN-2019): SIEMENS FURTHER INVESTIGATED THE ISSUE. THE PATIENT WAS REDRAWN AND THE RESULTS WERE HIGHER THAN EXPECTED. THE CUSTOMER RAN DILUTIONS ON THE PATIENT SAMPLE AND RESULTS OF THE DILUTIONS WERE ALSO HIGHER THAN EXPECTED. SIEMENS INSTRUCTED THE CUSTOMER TO CENTRIFUGE THE SAMPLE AND TEST THE PLASMA FOR TACROLIMUS (TAC). SIEMENS REQUESTED A LIST OF THE PATIENT'S MEDICATIONS. ADDITIONAL INFORMATION (03-JAN-2019): THE ORIGINAL PATIENT SAMPLE WAS TESTED BY AN ALTERNATE METHOD AND A LOW TAC RESULT WAS OBTAINED. SIEMENS ANALYZED THE INSTRUMENT DATA AND THERE WERE NO INSTRUMENT ISSUES SUCH AS CHROME OR FOAM. REVIEW OF THE FILTER DATA PROVIDED BY THE CUSTOMER SHOWED ONE SAMPLE ID FOR THIS PATIENT WAS RUN ONCE. SIEMENS INSTRUCTED THE CUSTOMER ON TROUBLESHOOTING WHEN THERE IS A SUSPECTED INTERFERENT FOR TAC. THE INFORMATION FOR USE (IFU) FOR TAC DOES NOT RECOMMEND PERFORMING DILUTIONS BEYOND 1:2. THE CUSTOMER DID NOT PROVIDE A LIST OF THE PATIENT'S MEDICATIONS. THIS EVENT IS ISOLATED TO A SINGLE PATIENT. THE RESULTS OF THE ALTERNATE TESTING AND THE DILUTION TESTING INDICATE THIS EVENT MAY BE DUE TO NONSPECIFIC BINDING (NSB). SIEMENS ALSO FILED SUPPLEMENTAL MDR# 2517506-2018-00681_S1 AND 2517506-2018-00682_S1 FOR THIS EVENT.
THE CUSTOMER CONTACTED A SIEMENS CUSTOMER CARE CENTER (CCC). QUALITY CONTROLS WERE WITHIN ACCEPTABLE RANGES. THE INSTRUMENT FILES WERE REVIEWED AND NO INSTRUMENT ISSUES WERE NOTED. SIEMENS HAS REVIEWED THE FILTER DATA FILES PROVIDED. THE DATA DID NOT SHOW ANY INSTRUMENT ISSUES SUCH AS CHROME, FOAM, OR OTHER ISSUES. THE CAUSE OF THE DISCORDANT, FALSELY ELEVATED TAC RESULTS IS UNKNOWN. SIEMENS IS INVESTIGATING THE ISSUE. SIEMENS ALSO FILED MDR# 2517506-2018-00674, 2517506-2018-00676, 2517506-2018-00677, 2517506-2018-00681 AND 2517506-2018-00682 FOR THIS EVENT.
DISCORDANT, FALSELY ELEVATED TACROLIMUS (TAC) RESULTS WERE OBTAINED ON A PATIENT SAMPLE ON A DIMENSION EXL WITH LM INSTRUMENT. THE INITIAL DISCORDANT RESULT WAS REPORTED TO THE PHYSICIAN(S) WHO QUESTIONED THE RESULT. THE SAMPLES WERE DILUTED AND REPEATED ON AN ALTERNATE DIMENSION EXL INSTRUMENT AND TWO NEW SAMPLE DRAWS WERE ALSO TESTED ON (B)(6) 2018 AND THE RESULTS WERE DISCORDANT. THE SAMPLE WAS TESTED BY AN ALTERNATE METHOD AND THE RESULTS WERE REPORTED TO THE PHYSICIAN(S) AS THE CORRECT RESULTS FOR THE PATIENT. THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED TAC RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 986452 | DIMENSION EXL WITH LM | CLINICAL CHEMISTRY ANALYZER | JJE | SIEMENS HEATLHCARE DIAGNOSTICS INC. | DIMENSION EXL WITH LM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |