FDA Adverse Event Malfunction Summary report: N

MO.MA ULTRA

MDR report key: 8141076 · Received December 7, 2018

Report

Report Number
3004066202-2018-00058
Event Type
Malfunction
Date Received
December 7, 2018
Date of Event
November 26, 2018
Report Date
February 11, 2019
Manufacturer
INVATEC SPA
Product Code
NTE
PMA / PMN Number
K092177
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

A KINK IN THE PLASTIC CHANNEL BETWEEN THE TWO CHANNELS WAS OBSERVED. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

EVALUATION SUMMARY: THE DEVICE WAS RETURNED WITHIN A SHIPPING BOX. THE DEVICE WAS ENVELOPED IN A PLASTIC POUCH, INSIDE A PAPER POUCH. THE INFORMATION REPORTED ON THE HUB WAS CONSISTENT WITH THE DATA IN THE COMPLAINT FORM RECEIVED (LOT# 1L000867). A TACTILE AND VISUAL INSPECTION WAS BEEN CONDUCTED: TRACES OF DRIED BLOOD WAS FOUND ON THE DEVICE. THE SHAFT WAS DAMAGED AND CUT IN THE DISTAL PART. ADDITIONAL INFORMATION: THE VESSEL WAS NOT PARTICULARLY TORTUOUS. NO IMAGES ARE AVAILABLE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

PHYSICIAN INTENDED TO USE A MOMA ULTRA DEVICE DURING TREATMENT. IFU WAS FOLLOWED AND THE DEVICE WAS PREPPED WITHOUT ISSUE. IT IS REPORTED THAT RESISTANCE WAS EXPERIENCED WHEN INSERTING THE 0.035¿ GUIDEWIRE INSIDE THE MOMA PRIOR TO INSERTING INTO THE PATIENT AND ALSO DURING THE FIRST PHASE OF THE DELIVERING OF THE SYSTEM INSIDE THE PATIENT. UNDER FLUOROSCOPY, AT THE LEVEL OF THE AORTA, IT WAS OBSERVED THAT THE ECA MARKER AND THE CCA MARKERS WERE CLOSE. THE DEVICE WAS REMOVED FROM THE PATIENT DUE TO THIS ISSUE. ON REMOVAL, A CUT IN THE PLASTIC CHANNEL BELOW THE ECA MARKER WAS OBSERVED. THE DEVICE WAS REPLACED WITH ANOTHER MOMA ULTRA. NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
981718 MO.MA ULTRA CATHETER, CAROTID, TEMPORARY, FOR EMBOLIZATION CAPTURE NTE INVATEC SPA

Patients

Seq Age Sex Outcome Treatment
1 71 YR