FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION S7 SYSTEM

MDR report key: 8140299 · Received December 6, 2018

Report

Report Number
1723170-2018-06108
Event Type
Malfunction
Date Received
December 6, 2018
Date of Event
November 9, 2018
Report Date
June 6, 2019
Manufacturer
MEDTRONIC NAVIGATION, INC
Product Code
HAW
PMA / PMN Number
K050438
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE SERIAL NUMBER AND UDI NUMBER UNAVAILABLE. NO PARTS HAVE BEEN RETURNED TO THE MANUFACTURER FOR ANALYSIS. DEVICE MANUFACTURING DATE IS UNAVAILABLE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

A MEDTRONIC REPRESENTATIVE (REP) WENT TO THE SITE TO TEST THE EQUIPMENT. THE REP CONNECTED THE SYSTEM'S CAMERA TO THE OTHER MEDTRONIC NAVIGATION SYSTEM ON SITE, AND THE CAMERA POWERED ON WITHOUT ISSUE. AS A RESULT, TWO POLARIS SPECTRA SYSTEM CONTROL UNIT (SCU) TO CAMERA COMMUNICATION CABLES WERE SENT TO THE SITE. THE REP REPLACED THE OLD SCU TO CAMERA COMMUNICATION CABLES WITH THESE NEW ONES. THE NAVIGATION SYSTEM THEN PASSED THE SYSTEM CHECKOUT AND WAS FOUND TO BE FULLY FUNCTIONAL. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE INTERNAL POLARIS SPECTRA SYSTEM CONTROL UNIT (SCU) TO POSITIONING SENSOR UNIT (PSU) CABLE WAS RETURNED TO MEDTRONIC FOR FURTHER EVALUATION. THE CABLE WAS RETURNED WITH THE FEMALE LEMO CONNECTOR END DISASSEMBLED. AFTER RE-ASSEMBLING THE CONNECTOR, THE CABLE PASSED A CONTINUITY TEST WITH NO OPENS OR SHORTS DETECTED. THERE WAS NO PROBLEM FOUND WITH THE INTERNAL SCU TO PSU CABLE. THE EXTERNAL SCU TO PSU CABLE WAS ALSO RETURNED TO MEDTRONIC FOR FURTHER EVALUATION. THE RETURNED CABLE HAD A BENT PIN IN THE STRAIGHT LEMO CONNECTOR. AS IS, THE CABLE WOULD NOT CONNECT TO THE MATING PART. AFTER STRAIGHTENING THE PIN, THE CABLE PASSED A CONTINUITY TEST WITH NO OPENS OR SHORTS DETECTED. ANALYSIS DETERMINED THERE WAS A PHYSICAL DAMAGE FAILURE WITH THE EXTERNAL SCU TO PSU CABLE. THIS ISSUE WAS DOCUMENTED IN A MEDTRONIC NAVIGATION HARDWARE ANOMALY TRACKING DATABASE. (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE NAVIGATION SYSTEM¿S CAMERA WAS NOT CONNECTED AFTER 30 MINUTES OF THE SYSTEM BEING ON. NO FAULT LIGHT WAS INDICATED ON THE CAMERA. IT WAS NOTED THAT THE CABLE DID NOT SEEM FRAYED. TO TROUBLESHOOT, A MEDTRONIC REPRESENTATIVE (REP) CHECKED THE NDI TOOLBOX, AND THE POSITIONING SENSOR UNIT (PSU) AND POLARIS SPECTRA SYSTEM CONTROL UNIT (SCU) DID NOT SHOW UP. THERE WERE GREEN LIGHTS ON THE CAMERA AND THE WIRES FROM THE CAMERA TO THE COMPUTER DID NOT LOOK DAMAGED. THE REP CHECKED THE SCU AND WHEN WIGGLING THE CABLE THAT CONNECTS FROM THE SCU TO THE EXTERNAL CABLE, THE REP WAS ABLE TO SEE THE PSU AND SCU IN THE NDI TOOLBOX FOR A MOMENT. THE ISSUE STILL PERSISTED, HOWEVER, AFTER BYPASSING THE INTERNAL CABLE. THE NDI TOOLBOX SHOWED THE SCU CONNECTED AND THE CAMERA NOT CONNECTED. THERE WERE NO FAULTS IN THE SCU. ON THE SCU THE LOWER LEFT LIGHT WAS SOLID GREEN AND THE RIGHT-SIDE LIGHTS WERE DOUBLE GREEN TO MATCH THE CAMERA, BUT EVERY 10-15 SECONDS THEY WOULD FLASH RED. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS OBSERVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
979689 STEALTHSTATION S7 SYSTEM INSTRUMENT, STEREOTAXIC HAW MEDTRONIC NAVIGATION, INC 9733856

Patients

Seq Age Sex Outcome Treatment
1