STEALTHSTATION S7 SYSTEM
Report
- Report Number
- 1723170-2018-06108
- Event Type
- Malfunction
- Date Received
- December 6, 2018
- Date of Event
- November 9, 2018
- Report Date
- June 6, 2019
- Manufacturer
- MEDTRONIC NAVIGATION, INC
- Product Code
- HAW
- PMA / PMN Number
- K050438
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
DEVICE SERIAL NUMBER AND UDI NUMBER UNAVAILABLE. NO PARTS HAVE BEEN RETURNED TO THE MANUFACTURER FOR ANALYSIS. DEVICE MANUFACTURING DATE IS UNAVAILABLE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
A MEDTRONIC REPRESENTATIVE (REP) WENT TO THE SITE TO TEST THE EQUIPMENT. THE REP CONNECTED THE SYSTEM'S CAMERA TO THE OTHER MEDTRONIC NAVIGATION SYSTEM ON SITE, AND THE CAMERA POWERED ON WITHOUT ISSUE. AS A RESULT, TWO POLARIS SPECTRA SYSTEM CONTROL UNIT (SCU) TO CAMERA COMMUNICATION CABLES WERE SENT TO THE SITE. THE REP REPLACED THE OLD SCU TO CAMERA COMMUNICATION CABLES WITH THESE NEW ONES. THE NAVIGATION SYSTEM THEN PASSED THE SYSTEM CHECKOUT AND WAS FOUND TO BE FULLY FUNCTIONAL. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
DEVICE EVALUATION: THE INTERNAL POLARIS SPECTRA SYSTEM CONTROL UNIT (SCU) TO POSITIONING SENSOR UNIT (PSU) CABLE WAS RETURNED TO MEDTRONIC FOR FURTHER EVALUATION. THE CABLE WAS RETURNED WITH THE FEMALE LEMO CONNECTOR END DISASSEMBLED. AFTER RE-ASSEMBLING THE CONNECTOR, THE CABLE PASSED A CONTINUITY TEST WITH NO OPENS OR SHORTS DETECTED. THERE WAS NO PROBLEM FOUND WITH THE INTERNAL SCU TO PSU CABLE. THE EXTERNAL SCU TO PSU CABLE WAS ALSO RETURNED TO MEDTRONIC FOR FURTHER EVALUATION. THE RETURNED CABLE HAD A BENT PIN IN THE STRAIGHT LEMO CONNECTOR. AS IS, THE CABLE WOULD NOT CONNECT TO THE MATING PART. AFTER STRAIGHTENING THE PIN, THE CABLE PASSED A CONTINUITY TEST WITH NO OPENS OR SHORTS DETECTED. ANALYSIS DETERMINED THERE WAS A PHYSICAL DAMAGE FAILURE WITH THE EXTERNAL SCU TO PSU CABLE. THIS ISSUE WAS DOCUMENTED IN A MEDTRONIC NAVIGATION HARDWARE ANOMALY TRACKING DATABASE. (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE NAVIGATION SYSTEM¿S CAMERA WAS NOT CONNECTED AFTER 30 MINUTES OF THE SYSTEM BEING ON. NO FAULT LIGHT WAS INDICATED ON THE CAMERA. IT WAS NOTED THAT THE CABLE DID NOT SEEM FRAYED. TO TROUBLESHOOT, A MEDTRONIC REPRESENTATIVE (REP) CHECKED THE NDI TOOLBOX, AND THE POSITIONING SENSOR UNIT (PSU) AND POLARIS SPECTRA SYSTEM CONTROL UNIT (SCU) DID NOT SHOW UP. THERE WERE GREEN LIGHTS ON THE CAMERA AND THE WIRES FROM THE CAMERA TO THE COMPUTER DID NOT LOOK DAMAGED. THE REP CHECKED THE SCU AND WHEN WIGGLING THE CABLE THAT CONNECTS FROM THE SCU TO THE EXTERNAL CABLE, THE REP WAS ABLE TO SEE THE PSU AND SCU IN THE NDI TOOLBOX FOR A MOMENT. THE ISSUE STILL PERSISTED, HOWEVER, AFTER BYPASSING THE INTERNAL CABLE. THE NDI TOOLBOX SHOWED THE SCU CONNECTED AND THE CAMERA NOT CONNECTED. THERE WERE NO FAULTS IN THE SCU. ON THE SCU THE LOWER LEFT LIGHT WAS SOLID GREEN AND THE RIGHT-SIDE LIGHTS WERE DOUBLE GREEN TO MATCH THE CAMERA, BUT EVERY 10-15 SECONDS THEY WOULD FLASH RED. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS OBSERVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 979689 | STEALTHSTATION S7 SYSTEM | INSTRUMENT, STEREOTAXIC | HAW | MEDTRONIC NAVIGATION, INC | 9733856 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |