FDA Adverse Event Injury Summary report: N

GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM

MDR report key: 8139916 · Received December 6, 2018

Report

Report Number
2024168-2018-09421
Event Type
Injury
Date Received
December 6, 2018
Date of Event
November 12, 2018
Report Date
February 13, 2019
Manufacturer
AV-TEMECULA-CT
Product Code
MAF
PMA / PMN Number
H000001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

INTERNAL FILE NUMBER - (B)(4). CORRECTION: LOT NUMBER HAS BEEN CHANGED FROM 8062541 TO UNKNOWN. THE DEVICE WAS NOT RETURNED FOR EVALUATION. A REVIEW OF THE LOT HISTORY RECORD AND COMPLAINT HISTORY OF THE REPORTED LOT COULD NOT BE CONDUCTED BECAUSE THE LOT NUMBER WAS NOT PROVIDED. THE INVESTIGATION DETERMINED THAT THE REPORTED FAILURE TO ADVANCE APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE OR LABELING OF THE DEVICE.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. IT IS UNKNOWN IF THE DEVICE IS RETURNING FOR ANALYSIS. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE OTHER GRAFTMASTER DEVICE REFERENCED IS FILED UNDER A SEPARATE MEDWATCH REPORT NUMBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A PERFORATION IN AN UNSPECIFIED CORONARY ARTERY. A 2.8 X 16 MM GRAFTMASTER COVERED STENT WAS IMPLANTED; HOWEVER, IT DID NOT SEAL THE PERFORATION COMPLETELY. THEREFORE, AN ATTEMPT WAS MADE TO ADVANCE A 2.8 X 26 MM GRAFTMASTER COVERED STENT; HOWEVER, THE DEVICE FAILED TO CROSS THE LESION AS IT COULD NOT CROSS THROUGH THE FIRST COVERED STENT. IT WAS REPORTED THAT THE PERFORATION WAS SEALED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
979003 GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM CORONARY STENT GRAFT MAF AV-TEMECULA-CT 8062541

Patients

Seq Age Sex Outcome Treatment
1 92 YR Required Intervention STENT: GRAFTMASTER 2.8X16.| STENT: GRAFTMASTER 2.8X16