GII QA+ O/C CP-2 *EA
Report
- Report Number
- 1221934-2018-55566
- Event Type
- Malfunction
- Date Received
- December 6, 2018
- Date of Event
- June 11, 2017
- Report Date
- June 11, 2017
- Manufacturer
- DEPUY MITEK LLC US
- Product Code
- JDR
- UDI-DI
- 10886705008612
- PMA / PMN Number
- K051989
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER
Narratives
UDI: (B)(4)-INCOMPLETE. THE LOT NUMBER IS CURRENTLY UNAVAILABLE; THEREFORE, THE EXP DATE IS UNAVAILABLE. INVESTIGATION SUMMARY: P/N: 222983: THE COMPLAINT DEVICE IS NOT AVAILABLE FOR A PHYSICAL EVALUATION. NO FURTHER INFORMATION REGARDING THE TECHNIQUE OR INSTRUMENTS USED HAS BEEN PROVIDED TO DETERMINE A ROOT CAUSE FOR THIS FAILURE. FURTHERMORE, NO LOT NUMBERS WERE SUPPLIED WHICH PRECLUDES CONDUCTING A DHR REVIEW OR A LOT SPECIFIC SEARCH IN THE COMPLAINTS HANDLING SYSTEM. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, SHOULD ANY NEW INFORMATION BE PROVIDED IN FUTURE, THIS FILE WILL BE RE-OPENED AND A THOROUGH INVESTIGATION WILL BE PERFORMED. DEPUY MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.
THIS IS REPORT 2 OF 2 FOR THE SAME EVENT. IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE GII QUICKANCHOR ORTHOCORD DEVICE AND GRYPHON P BR DS ANCHOR W/OC DEVICE GOT ENTANGLED IN THE PATIENT'S TISSUES AND DISENGAGED WHILE TRYING TO PUT THE ANCHOR AND BROKE. IT WAS NOT REPORTED IF THERE WAS A DELAY IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 977032 | GII QA+ O/C CP-2 *EA | SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE | JDR | DEPUY MITEK LLC US | UNK | 10886705008612 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |