FDA Adverse Event Malfunction Summary report: N

MAGNA PURE LC INSTRUMENT

MDR report key: 813522 · Received November 12, 2004

Report

Report Number
1823260-2004-01297
Event Type
Malfunction
Date Received
November 12, 2004
Date of Event
October 26, 2004
Report Date
October 26, 2004
Manufacturer
PRECISION SYSTEMS SCIENCES
Product Code
JJE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

ACCOUNT HAS NOTICED THAT COLD STORAGE BLOCK AND POST ELUTION BLOCK ARE AT ROOM TEMPERATURE WHEN A RUN IS COMPLETE. A SOFTWARE BUG HAS BEEN IDENTIFIED WITH SOFTWARE VERSION 3.09 WHERE THE COOLING BLOCKS SHUT DOWN AFTER A UV DECONTAMINATION; LEAKAGE TEST OF Z-LEVEL ALIGNMENT AND DO NOT TURN BACK ON. HOWEVER, THE SYSTEM DISPLAYS THE COOLING BLOCK STATUS AS "PASS" INCORRECTLY INDICATING THAT THE COOLING BLOCKS ARE AT THE APPROPRIATE TEMPERATURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAGNA PURE LC INSTRUMENT GENERAL PURPOSE INSTRUMENT JJE PRECISION SYSTEMS SCIENCES NA NA