FDA Adverse Event Malfunction Summary report: N

HANDPIECE MICS

MDR report key: 8134974 · Received December 5, 2018

Report

Report Number
3005985723-2018-00715
Event Type
Malfunction
Date Received
December 5, 2018
Date of Event
November 26, 2018
Report Date
August 6, 2019
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486030193
PMA / PMN Number
K142530
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

FOLLOW-UP #1 AND FINAL REPORT SUBMITTED TO UPDATE SECTIONS D.3, G.1, G.4, G.7, H.2, H.3, H.6, H.10 AND H.11 BASED ON THE RESULTS OF INVESTIGATION. REPORTED EVENT: IT WAS REPORTED THAT WHEN SWAPPING ATTACHMENTS AN INNER RING METAL PIECE FELL OUT FROM INSIDE OF MICS. SWAPPED OUT MICS. PRODUCT EVALUATION AND RESULTS: PRODUCT INSPECTION COULD NOT BE PERFORMED AS THE PRODUCT WAS NOT AVAILABLE FOR EVALUATION. PRODUCT HISTORY REVIEW: DEVICE HISTORY RECORDS INDICATE 25 DEVICES WERE MANUFACTURED UNDER LOT K0B9X AND 23 DEVICES WERE ACCEPTED INTO FINAL STOCK ON 8/25/2017. A REVIEW OF QT 18-05-0071 REVEALED THAT THE ISSUE IS NOT RELATED TO THE FAILURE ALLEGED IN THIS COMPLIANT. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS RELATED TO P/N 209063, PRODEX LOT K0B9X SHOWS 02 ADDITIONAL COMPLAINT(S) RELATED TO THE FAILURE IN THIS INVESTIGATION.COMPLAINT ID:(B)(4). CONCLUSIONS: THE FAILURE MODE COULD NOT BE CONFIRMED BECAUSE THE PART WAS NOT AVAILABLE FOR EVALUATION. IF DEVICE AND/OR ADDITIONAL INFORMATION BECOME AVAILABLE, THIS INVESTIGATION WILL BE REOPENED. CORRECTIVE ACTION/PREVENTIVE ACTION: A REVIEW OF STRYKER¿S NC/CAPA DATABASE INDICATED THERE HAVE BEEN AN NC AND CAPA ASSOCIATED WITH THE PRODUCT AND FAILURE MODE REPORTED IN THIS EVENT. THIS IS NC1414517 AND CAPA (B)(4).

Description of Event or Problem · 0

WHEN SWAPPING ATTACHMENTS AN INNER RING METAL PIECE FELL OUT FROM INSIDE OF MICS. SWAPPED OUT MICS. CASE TYPE: TKA SURGICAL DELAY UNDER FIVE MINUTES.

Additional Manufacturer Narrative · 1

¿AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.¿

Description of Event or Problem · 1

WHEN SWAPPING ATTACHMENTS AN INNER RING METAL PIECE FELL OUT FROM INSIDE OF MICS. SWAPPED OUT MICS. CASE TYPE: TKA. SURGICAL DELAY UNDER FIVE MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
974911 HANDPIECE MICS STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 42020518 /4204027 00848486030193

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization