HANDPIECE MICS
Report
- Report Number
- 3005985723-2018-00715
- Event Type
- Malfunction
- Date Received
- December 5, 2018
- Date of Event
- November 26, 2018
- Report Date
- August 6, 2019
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- UDI-DI
- 00848486030193
- PMA / PMN Number
- K142530
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 AND FINAL REPORT SUBMITTED TO UPDATE SECTIONS D.3, G.1, G.4, G.7, H.2, H.3, H.6, H.10 AND H.11 BASED ON THE RESULTS OF INVESTIGATION. REPORTED EVENT: IT WAS REPORTED THAT WHEN SWAPPING ATTACHMENTS AN INNER RING METAL PIECE FELL OUT FROM INSIDE OF MICS. SWAPPED OUT MICS. PRODUCT EVALUATION AND RESULTS: PRODUCT INSPECTION COULD NOT BE PERFORMED AS THE PRODUCT WAS NOT AVAILABLE FOR EVALUATION. PRODUCT HISTORY REVIEW: DEVICE HISTORY RECORDS INDICATE 25 DEVICES WERE MANUFACTURED UNDER LOT K0B9X AND 23 DEVICES WERE ACCEPTED INTO FINAL STOCK ON 8/25/2017. A REVIEW OF QT 18-05-0071 REVEALED THAT THE ISSUE IS NOT RELATED TO THE FAILURE ALLEGED IN THIS COMPLIANT. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS RELATED TO P/N 209063, PRODEX LOT K0B9X SHOWS 02 ADDITIONAL COMPLAINT(S) RELATED TO THE FAILURE IN THIS INVESTIGATION.COMPLAINT ID:(B)(4). CONCLUSIONS: THE FAILURE MODE COULD NOT BE CONFIRMED BECAUSE THE PART WAS NOT AVAILABLE FOR EVALUATION. IF DEVICE AND/OR ADDITIONAL INFORMATION BECOME AVAILABLE, THIS INVESTIGATION WILL BE REOPENED. CORRECTIVE ACTION/PREVENTIVE ACTION: A REVIEW OF STRYKER¿S NC/CAPA DATABASE INDICATED THERE HAVE BEEN AN NC AND CAPA ASSOCIATED WITH THE PRODUCT AND FAILURE MODE REPORTED IN THIS EVENT. THIS IS NC1414517 AND CAPA (B)(4).
WHEN SWAPPING ATTACHMENTS AN INNER RING METAL PIECE FELL OUT FROM INSIDE OF MICS. SWAPPED OUT MICS. CASE TYPE: TKA SURGICAL DELAY UNDER FIVE MINUTES.
¿AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.¿
WHEN SWAPPING ATTACHMENTS AN INNER RING METAL PIECE FELL OUT FROM INSIDE OF MICS. SWAPPED OUT MICS. CASE TYPE: TKA. SURGICAL DELAY UNDER FIVE MINUTES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 974911 | HANDPIECE MICS | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. | 42020518 /4204027 | 00848486030193 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |