FREESTYLE LITE
Report
- Report Number
- 2954323-2018-09309
- Event Type
- Malfunction
- Date Received
- December 5, 2018
- Date of Event
- November 15, 2018
- Report Date
- January 10, 2019
- Manufacturer
- ABBOTT DIABETES CARE INC.
- Product Code
- NBW
- PMA / PMN Number
- K092602
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- 003
Narratives
THE PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW- UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. NOTE: THE METER IS DESIGNED TO REPORT READINGS OF 20 MG/DL TO 500 MG/DL. IT SHOULD BE NOTED THAT THIS METER DOES NOT GIVE NUMERIC VALUE FOR READINGS LESS THAN 20 MG/DL. A "LO" DISPLAY MESSAGE INDICATES A READING LESS THAN 20 MG/DL. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
ADDITIONAL INFORMATION: (STRIP LOT NUMBER) AND (EXPIRATION DATE) HAS BEEN UPDATED BASED ON THE RETURNED PRODUCT. THE REPORTED METER AND STRIP LOT NUMBER 1060316 HAVE BEEN RETURNED AND INVESTIGATED. VISUAL INSPECTION HAS BEEN PERFORMED ON THE REPORTED METER AND NO ISSUES WERE OBSERVED. THE METER POWERED ON WITH BUTTON DEPRESSION AND WITH STRIP INSERTION. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE REPORTED READINGS WERE NOT FOUND IN THE METER'S MEMORY.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 130 MG/DL, 20 MG/DL, 140 MG/DL AND 30 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 130 MG/DL, 20 MG/DL, 140 MG/DL AND 30 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 973083 | FREESTYLE LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | ABBOTT DIABETES CARE INC. | 71501-01 | 1060316 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |