FDA Adverse Event Injury Summary report: N

TITAN TOUCH SCRO ZERO ANG 20CM

MDR report key: 8130694 · Received December 4, 2018

Report

Report Number
2125050-2018-00863
Event Type
Injury
Date Received
December 4, 2018
Date of Event
October 16, 2018
Report Date
February 28, 2019
Manufacturer
COLOPLAST A/S
Product Code
FHW
UDI-DI
05708932539197
PMA / PMN Number
P000006
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP MDR IS CREATED TO DOCUMENT THE EVALUATION OF THE RETURNED DEVICE. A TITAN TOUCH PUMP, TWO CYLINDERS, AND A RESERVOIR WERE RECEIVED FOR EVALUATION. EXAMINATION AND TESTING OF THE RETURNED COMPONENTS REVEALED SURFACE ABRASION ON ALL TUBES OF THE PUMP, CYLINDER #1 AND CYLINDER #2. NO FUNCTIONAL ABNORMALITIES ARE NOTED WITH THE PUMP, CYLINDER #1, CYLINDER #2, OR THE RESERVOIR. THE INFORMATION RECEIVED INDICATED LOSS OF FLUID, BUT BECAUSE NO FUNCTIONAL ABNORMALITIES WERE NOTED WITH THE RETURNED COMPONENTS, QUALITY CANNOT CONFIRM THE COMPLAINT AS REPORTED. QUALITY REVIEWED THE MANUFACTURING RECORDS FOR THIS DEVICE AND CONFIRMED THAT THERE WERE NO DISCREPANCIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT AND VERIFIED THAT THE DEVICES FROM THIS LOT MET ALL SPECIFICATIONS PRIOR TO RELEASE. A REVIEW OF THE COMPLAINT HISTORY DATABASE REVEALED NO SIGNIFICANT TRENDS FOR LOT 6203704. REVIEW OF NONCONFORMING REPORTS REVEALED NO NONCONFORMANCE'S FOR THIS LOT. NO CAPAS ARE ASSOCIATED WITH THIS LOT.

Additional Manufacturer Narrative · 1

WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCURRENCE. ONCE OUR EVALUATION IS COMPLETE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

ACCORDING TO THE AVAILABLE INFORMATION, LOSS OF FLUID; PINHOLE PUNCTURE OBSERVED DURING EXPLANT UPON FILLING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
970717 TITAN TOUCH SCRO ZERO ANG 20CM INFLATABLE PENILE PROSTHESIS FHW COLOPLAST A/S ES29202400 6203704 05708932539197

Patients

Seq Age Sex Outcome Treatment
1 72 YR Other