FDA Adverse Event Malfunction Summary report: N

EVIS EXERA LLL COLONOVIDEOSCOPE

MDR report key: 8127357 · Received December 3, 2018

Report

Report Number
8010047-2018-02354
Event Type
Malfunction
Date Received
December 3, 2018
Date of Event
October 16, 2018
Report Date
January 16, 2019
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FDF
UDI-DI
04953170305177
PMA / PMN Number
K112680
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) BUT WAS RETURNED TO OLYMPUS (B)(4). (B)(4) SENT THE SUBJECT DEVICE TO A THIRD PARTY LABORATORY FOR ADDITIONAL MICROBIOLOGICAL TESTING. AS A RESULT OF THE TESTING, NO MICROBE WAS DETECTED FROM THE SAMPLE COLLECTED FROM ALL THE CHANNELS OF THE SUBJECT DEVICE. THE TESTING RESULT CLEARED THE (B)(4) GUIDELINE. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED.

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE IN THIS REPORT HAS NOT BEEN RETURNED TO OMSC FOR EVALUATION. OMSC REVIEWED THE MANUFACTURING HISTORY OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF ROUTINE MICROBIOLOGICAL TESTING BY THE USER FACILITY, THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE TESTED POSITIVE FOR GRAM NEGATIVE BACILLUS OF ENTEROBACTER GENUS AND PSEUDOMONAS SPP (OVER 100 CFU / ENDOSCOPE). THE DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, SOLUSCOPE S4, USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
962895 EVIS EXERA LLL COLONOVIDEOSCOPE COLONOVIDEOSCOPE FDF OLYMPUS MEDICAL SYSTEMS CORP. CF-H190I 04953170305177

Patients

Seq Age Sex Outcome Treatment
1