FDA Adverse Event Malfunction Summary report: N

2.7 DEGREE STRAIGHT SAGITAL SAW

MDR report key: 8127047 · Received December 3, 2018

Report

Report Number
3005985723-2018-00702
Event Type
Malfunction
Date Received
December 3, 2018
Date of Event
November 6, 2018
Report Date
March 25, 2019
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486032111
PMA / PMN Number
K143752
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

REPORTED EVENT: IT WAS REPORTED THAT MICS ATTACHMENT NUT BROKE. PRODUCT EVALUATION AND RESULTS: NO DEVICE INSPECTION COULD BE COMPLETED AS THE PRODUCT WAS NOT AVAILABLE FOR EVALUATION. ALTHOUGH A PICTURE OF THE DEVICE WAS PROVIDED BUT NOTHING RELEVANT TO THE CURRENT EVENT COULD BE DETERMINED FROM IT. PRODUCT HISTORY REVIEW: REVIEW OF THE PRODUCT HISTORY RECORDS INDICATE (B)(4) DEVICES WERE MANUFACTURED UNDER LOT NO 35030117 AND 43 DEVICES WERE ACCEPTED INTO FINAL STOCK ON 03/06/2017. REVIEW OF QT17-03-0016 REVEALED THAT THE NON-CONFORMANCE IS NOT RELATED TO THE FAILURE ALLEGED IN THIS COMPLIANT. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 212186, LOT NUMBER 35030117, SHOWS NO ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSIONS: THE FAILURE MODE COULD NOT BE CONFIRMED BECAUSE THE PART WAS NOT AVAILABLE FOR EVALUATION. IF DEVICE AND/OR ADDITIONAL INFORMATION BECOME AVAILABLE, THIS INVESTIGATION WILL BE REOPENED. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.

Description of Event or Problem · 0

TKA. MICS ATTACHMENT NUT BROKE.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

TKA. MICS ATTACHMENT NUT BROKE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
964010 2.7 DEGREE STRAIGHT SAGITAL SAW STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 35030117 00848486032111

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization