FDA Adverse Event Malfunction Summary report: N

VITROS CHEMISTRY PRODUCTS NA+ SLIDES

MDR report key: 8126090 · Received December 3, 2018

Report

Report Number
1319809-2018-00145
Event Type
Malfunction
Date Received
December 3, 2018
Date of Event
October 1, 2018
Report Date
December 3, 2018
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
JGS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONCLUDED THAT LOWER THAN EXPECTED VITROS CHEMISTRY PRODUCTS SODIUM (NA+) RESULTS WERE ATTAINED FROM COLLEGE OF AMERICAN PATHOLOGISTS (CAP) PROFICIENCY SAMPLES USING VITROS NA+ SLIDES IN COMBINATION WITH A VITROS 350 CHEMISTRY SYSTEM. THE INVESTIGATION COULD NOT DETERMINE AN ASSIGNABLE CAUSE OF THE EVENT. A REAGENT ISSUE AND/OR A CALIBRATION ISSUE CANNOT BE RULED OUT AS CONTRIBUTING FACTORS AS QUALITY CONTROL RESULTS FROM THE TIME OF THE EVENT WERE NOT MEETING EXPECTATIONS. CONTINUAL TRACKING AND TRENDING OF COMPLAINTS HAS NOT IDENTIFIED ANY SIGNALS THAT WOULD POINT TO A POTENTIAL SYSTEMIC ISSUE WITH VITROS NA+ SLIDES LOT: 4220-0994-1710. THE ISSUE WAS LATER RESOLVED WHEN AN ALTERNATE REAGENT LOT WAS CALIBRATED AND PUT INTO USE, AND THE SAMPLES WERE TESTED AGAIN. IT IS UNKNOWN IF THE INSTRUMENT WAS PERFORMING AS EXPECTED AT THE TIME OF THE EVENT AS PERFORMANCE TESTING WAS NOT CONDUCTED DURING THAT TIME. THEREFORE, THE INSTRUMENT CANNOT BE COMPLETELY RULED OUT AS A CONTRIBUTING FACTOR.

Description of Event or Problem · 1

A CUSTOMER REPORTED A LOWER THAN EXPECTED VITROS SODIUM (NA+) RESULT WAS OBTAINED FROM A COLLEGE OF AMERICAN PATHOLOGISTS (CAP) PROFICIENCY SAMPLE USING VITROS CHEMISTRY PRODUCTS SODIUM (NA+) SLIDES ON A VITROS 350 CHEMISTRY SYSTEM. COLLEGE OF AMERICAN PATHOLOGISTS C-C 2018 SAMPLE CHM-13: NA+ RESULT OF 144 MMOL/L VERSUS AN EXPECTED RESULT OF 150.3 MMOL/L. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER THAN EXPECTED VITROS NA+ RESULT WAS OBTAINED WHEN PROCESSING CAP FLUIDS WITH NO PATIENT RESULTS BEING REPORTED OUTSIDE OF THE LABORATORY. THERE WERE NO ALLEGATIONS OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. COMPLAINT NUMBER (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
966512 VITROS CHEMISTRY PRODUCTS NA+ SLIDES IN VITRO DIAGNOSTICS JGS ORTHO-CLINICAL DIAGNOSTICS 4220-0994-1710

Patients

Seq Age Sex Outcome Treatment
1