VITROS CHEMISTRY PRODUCTS NA+ SLIDES
Report
- Report Number
- 1319809-2018-00145
- Event Type
- Malfunction
- Date Received
- December 3, 2018
- Date of Event
- October 1, 2018
- Report Date
- December 3, 2018
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- JGS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION CONCLUDED THAT LOWER THAN EXPECTED VITROS CHEMISTRY PRODUCTS SODIUM (NA+) RESULTS WERE ATTAINED FROM COLLEGE OF AMERICAN PATHOLOGISTS (CAP) PROFICIENCY SAMPLES USING VITROS NA+ SLIDES IN COMBINATION WITH A VITROS 350 CHEMISTRY SYSTEM. THE INVESTIGATION COULD NOT DETERMINE AN ASSIGNABLE CAUSE OF THE EVENT. A REAGENT ISSUE AND/OR A CALIBRATION ISSUE CANNOT BE RULED OUT AS CONTRIBUTING FACTORS AS QUALITY CONTROL RESULTS FROM THE TIME OF THE EVENT WERE NOT MEETING EXPECTATIONS. CONTINUAL TRACKING AND TRENDING OF COMPLAINTS HAS NOT IDENTIFIED ANY SIGNALS THAT WOULD POINT TO A POTENTIAL SYSTEMIC ISSUE WITH VITROS NA+ SLIDES LOT: 4220-0994-1710. THE ISSUE WAS LATER RESOLVED WHEN AN ALTERNATE REAGENT LOT WAS CALIBRATED AND PUT INTO USE, AND THE SAMPLES WERE TESTED AGAIN. IT IS UNKNOWN IF THE INSTRUMENT WAS PERFORMING AS EXPECTED AT THE TIME OF THE EVENT AS PERFORMANCE TESTING WAS NOT CONDUCTED DURING THAT TIME. THEREFORE, THE INSTRUMENT CANNOT BE COMPLETELY RULED OUT AS A CONTRIBUTING FACTOR.
A CUSTOMER REPORTED A LOWER THAN EXPECTED VITROS SODIUM (NA+) RESULT WAS OBTAINED FROM A COLLEGE OF AMERICAN PATHOLOGISTS (CAP) PROFICIENCY SAMPLE USING VITROS CHEMISTRY PRODUCTS SODIUM (NA+) SLIDES ON A VITROS 350 CHEMISTRY SYSTEM. COLLEGE OF AMERICAN PATHOLOGISTS C-C 2018 SAMPLE CHM-13: NA+ RESULT OF 144 MMOL/L VERSUS AN EXPECTED RESULT OF 150.3 MMOL/L. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER THAN EXPECTED VITROS NA+ RESULT WAS OBTAINED WHEN PROCESSING CAP FLUIDS WITH NO PATIENT RESULTS BEING REPORTED OUTSIDE OF THE LABORATORY. THERE WERE NO ALLEGATIONS OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. COMPLAINT NUMBER (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 966512 | VITROS CHEMISTRY PRODUCTS NA+ SLIDES | IN VITRO DIAGNOSTICS | JGS | ORTHO-CLINICAL DIAGNOSTICS | 4220-0994-1710 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |