CERETOM
Report
- Report Number
- 3004938766-2007-00001
- Event Type
- Other
- Date Received
- January 24, 2007
- Date of Event
- December 22, 2006
- Report Date
- January 23, 2007
- Manufacturer
- NEUROLOGICA CORPORATION
- Product Code
- JAK
- PMA / PMN Number
- K051765
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INITIAL TEST BY NEUROLOGICA ENGINEER SHOWED UNIT WAS FUNCTIONING PROPERLY. THE INITIAL VIEWING OF THE IMAGES SHOWED WHAT APPEARED TO BE AN ARTIFACT CAUSED BY THE IMPROPER POSITIONING OF THE PATIENT'S HEAD IN THE SCANNER AND A METALLIC OBJECT IMPLEMENTED LESS THAN 5 MM FROM THE IMAGE PLAN. THE PATIENT'S HEAD WAS POSITIONED INCORRECTLY (NOT CENTERED, ANGLED) ACCORDING TO THE CERETOM USER MANUAL AND TRAINING PREVIOUSLY PROVIDED BY NEUROLOGICA. IT IS WELL KNOW IN CT IMAGING THAT HEAD SCANS PERFORMED OFF-CENTER CAN PRODUCE ARTIFACTS. AN IMPLANTED METAL ARTIFACT IS KNOWN IN CT IMAGING TO AFFECT THE CT-VALUE IN THE VICINITY OF THE IMPLANT. THE EFFECT DEPENDS ON THE LOCATION, SIZE AND DENSITY OF THE METAL OBJECT. FOLLOW-UP: THERE WAS NO PATIENT INJURY TO REPORT. THE HOSPITAL AND TECHNICIAN WERE REINSTRUCTED BY NEUROLOGICA TO POSITION PATIENTS ACCORDING TO INSTRUCTIONS IN THE USER MANUAL A TYPICAL CERETOM PATIENT SCAN COVERS THE DISTANCE FROM THE NOSE TO THE TOP OF THE SKULL, APPROXIMATELY 16 CM. THE INITIAL VIEWING OF THE IMAGES SHOWED WHAT APPEARED TO BE AN ARTIFACT CAUSED BY THE IMPROPER POSITIONING OF THE PATIENT'S HEAD IN THE SCANNER AND A METALLIC OBJECT IMPLEMENTED LESS THAN 5 MM FROM THE IMAGE PLAN. IN GENERAL, IN A FIXED SCANNER, THE PATIENT'S HEAD POSITION IS RESTRAINED BY THE SCANNER TABLE AND A HEAD HOLDER. HOWEVER, POSITIONING OF A PATIENT'S HEAD IN THE CERETOM DEPENDS UPON THE PATIENT'S CONDITION SINCE HE/SHE IS SCANNED ON HIS OWN BED. WITH THE PATIENT'S HEAD TILTED IN THE Y-DIRECTION, THE RIGHT HEMISPHERE LOOKED BIGGER AS THE SCAN TRAVEL TOWARD THE TOP OF THE SKULL. THE TILT CAUSED ASYMMETRY IN THE IMAGE WHICH AFFECTED THE BONE CORRECTION ON THE LEFT SIDE OF THE IMAGE. AN IMPLANTED METAL ARTIFACT IS KNOWN IN CT IMAGING TO AFFECT THE CT-VALUE IN THE VICINITY OF THE IMPLANT. THE EFFECT DEPENDS ON THE LOCATION, SIZE AND DENSITY OF THE METAL OBJECT. THE CERETOM USER MANUAL PROVIDES WRITTEN INSTRUCTION ON PROPER HEAD POSITIONING. THE POSITION OF THE SCAN NEAR THE TOP OF THE SKULL CAN HAVE A GREAT INFLUENCE ON THE CORRECTION DUE TO THE LARGE GRADIENT OF THE SKULL.
A PATIENT SCAN WAS MISDIAGNOSED AS A STROKE. AS A RESULT, THE DOCTOR REQUESTED A FOLLOW UP SCAN ON A FIXED SCANNER WHICH SHOWED NO SIGN OF A STROKE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CERETOM | CT SCANNER | JAK | NEUROLOGICA CORPORATION | NL3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |