FDA Adverse Event Malfunction Summary report: N

ION TORRENT¿ ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY

MDR report key: 8122601 · Received December 1, 2018

Report

Report Number
3003335080-2018-00002
Event Type
Malfunction
Date Received
December 1, 2018
Date of Event
October 29, 2018
Report Date
November 30, 2018
Manufacturer
LIFE TECHNOLOGIES CORPORATION
Product Code
OTV
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE ION TORRENT¿ ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY IS LABELED "FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES." (RUO) NO PATIENT IMPACT AT THE BIOCEPT SITE WAS REPORTED, HOWEVER, THE MANUFACTURER, THERMO FISHER SCIENTIFIC, IS SUBMITTING THIS MDR DUE TO A POSSIBLE FALSE POSITIVE RESULT` FOR THE T790M VARIANT AT THE (B)(6) IF THE LAB UTILIZES THE RUO PRODUCT TO ASSESS PATIENT SAMPLES. THE UNIT IN QUESTION WAS NOT RETURNED BY THE CUSTOMER. HOWEVER, A DIFFERENT UNIT FROM THE SAME LOT WAS INTERNALLY INVESTIGATED BY THE MANUFACTURER. THIS INVESTIGATION IS ONGOING. THERMO FISHER SCIENTIFIC HAS INITIATED THE RECALL PROCESS TO REMOVE UNITS OF ION TORRENT¿ ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY LOT #: 2025404 FROM THE MARKET. AFFECTED CUSTOMERS HAVE BEEN CONTACTED TO STOP USE OF THE PRODUCT.

Description of Event or Problem · 1

A REPRESENTATIVE FROM (B)(6), MET WITH A CUSTOMER REPRESENTATIVE FROM (B)(4) TO DISCUSS TECHNICAL ISSUES REGARDING VALIDATION ACTIVITIES AT THE (B)(6). (B)(6) COMPLAINED THAT A POSSIBLE CONTAMINANT IN THE ION TORRENT¿ ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY CAUSED THE GENETIC VARIANT, T790M, TO SHOW POSITIVE IN SAMPLE KNOWN TO BE NEGATIVE FOR THE VARIANT. THIS WAS HAULTING THE SITE'S VALIDATION ACTIVITIES. NO PATIENT SAMPLES WERE BEING ANALYZED WHEN THE CONTAMINATION WAS REPORTED. THE PRODUCT WAS NOT USED TO DETERMINE TREATMENT OR DIAGNOSTIC DECISIONS. ION TORRENT¿ ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY IS LABELED "FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES." THE PRODUCT IS NOT A CLEARED OR APPROVED MEDICAL DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
962591 ION TORRENT¿ ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY OTV LIFE TECHNOLOGIES CORPORATION 2025404

Patients

Seq Age Sex Outcome Treatment
1