FDA Adverse Event Malfunction Summary report: N

LIBERTY SELECT CYCLER ASSY(NON-VALUATED)

MDR report key: 8121127 · Received November 30, 2018

Report

Report Number
2937457-2018-03523
Event Type
Malfunction
Date Received
November 30, 2018
Date of Event
November 16, 2018
Report Date
December 26, 2018
Manufacturer
CONCORD MANUFACTURING
Product Code
FKX
UDI-DI
00840861102068
PMA / PMN Number
K181108
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

UPDATED 510(K) NUMBER TO K181108. PLANT INVESTIGATION: THE ACTUAL DEVICE WAS RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION. A VISUAL INSPECTION OF THE RETURNED CYCLER EXTERIOR SHOWED NO SIGN OF PHYSICAL DAMAGE. THERE ARE VISUAL INDICATIONS OF DRIED FLUID WITHIN THE CASSETTE COMPARTMENT. THERE WERE NO BURRS OR SHARP EDGES IN CASSETTE AREA THAT MAY HAVE PUNCTURED A CASSETTE MEMBRANE. THE VALVE ACTUATION TEST PASSED. THE SYSTEM AIR LEAK TEST PASSED. A POST-AST 2 HOUR 15 MIN 8500 ML SIMULATED TREATMENT WAS PERFORMED AND COMPLETED. A DC (DRAIN COMPLICATION ENCOUNTERED) SUPPORT WARNING OCCURRED, AS EXPECTED, WHEN THE PATIENT LINE WAS CLAMPED DURING THE DWELL PHASE CYCLE 1. THE PATIENT SENSOR CALIBRATION CHECK PASSED. THE MUSHROOM HEAD CHECK PASSED. THE CYCLER TESTED POSITIVE FOR GLUCOSE. AN INTERNAL VISUAL INSPECTION OF THE CYCLER FOUND VISUAL EVIDENCE OF DRIED FLUID UNDER PUMP ¿A¿ AND ¿B¿ MUSHROOM HEAD AND WITHIN THE RECESS OF THE BOTTOM COVER ADJACENT TO THE PUMP. THE CAUSE OF THE OBSERVED DRIED FLUID AND FLUID COULD NOT BE DETERMINED. A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NON-CONFORMANCES DURING THE MANUFACTURING PROCESS. IN ADDITION, A DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED AND VERIFIED THAT THE RESULTS OF THE IN-PROGRESS AND FINAL QUALITY CONTROL (QC) TESTING MET ALL REQUIREMENTS. UPON COMPLETION OF THE EVALUATION, THERE WERE NO MALFUNCTIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE CYCLER PERFORMED AS DESIGNED AND AN ASSOCIATED CAUSE COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 1

THE PLANT INVESTIGATION IS IN PROCESS. A SUPPLEMENTAL MDR WILL BE SUBMITTED UPON COMPLETION OF THIS ACTIVITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PERITONEAL DIALYSIS (PD) PATIENT DISCOVERED A FLUID LEAK ON THE INSIDE OF THE CASSETTE DOOR OF THEIR CYCLER DURING THEIR PD TREATMENT ON 11/16/2018. THE PATIENT WAS ADVISED TO DISCONTINUE THE USE OF THEIR CYCLER. UPON FOLLOW UP, THE PERITONEAL DIALYSIS NURSE (PDRN) STATED THE PATIENT IS TRAINED ON PERFORMING STAT DRAINS, AS WELL AS MANUAL PERITONEAL DIALYSIS THERAPY IF NEEDED, AND OPTED NOT TO COMPLETE THEIR TREATMENT ON (B)(6) 2018. THE PATIENT DID NOT DEVELOP ANY SYMPTOMS, ADVERSE EVENTS, INJURIES, OR REQUIRE MEDICAL INTERVENTION AS A RESULT OF THE REPORTED EVENT. THE PATIENT HAS RECEIVED A NEW CYCLER, WHICH IS WORKING WELL AND IS CONTINUING PERITONEAL DIALYSIS THERAPY WITH NO FURTHER ISSUES. THE CASSETTE USED BY THE PATIENT WAS DISCARDED AND IS NOT AVAILABLE FOR RETURN FOR PHYSICAL EVALUATION BY THE MANUFACTURER. THE CYCLER WAS RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
959021 LIBERTY SELECT CYCLER ASSY(NON-VALUATED) SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX CONCORD MANUFACTURING 00840861102068

Patients

Seq Age Sex Outcome Treatment
1 54 YR DELFLEX PD FLUID| LIBERTY CYCLER SET| DELFLEX PD FLUID| LIBERTY CYCLER SET