LIBERTY SELECT CYCLER ASSY(NON-VALUATED)
Report
- Report Number
- 2937457-2018-03523
- Event Type
- Malfunction
- Date Received
- November 30, 2018
- Date of Event
- November 16, 2018
- Report Date
- December 26, 2018
- Manufacturer
- CONCORD MANUFACTURING
- Product Code
- FKX
- UDI-DI
- 00840861102068
- PMA / PMN Number
- K181108
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- 003
Narratives
UPDATED 510(K) NUMBER TO K181108. PLANT INVESTIGATION: THE ACTUAL DEVICE WAS RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION. A VISUAL INSPECTION OF THE RETURNED CYCLER EXTERIOR SHOWED NO SIGN OF PHYSICAL DAMAGE. THERE ARE VISUAL INDICATIONS OF DRIED FLUID WITHIN THE CASSETTE COMPARTMENT. THERE WERE NO BURRS OR SHARP EDGES IN CASSETTE AREA THAT MAY HAVE PUNCTURED A CASSETTE MEMBRANE. THE VALVE ACTUATION TEST PASSED. THE SYSTEM AIR LEAK TEST PASSED. A POST-AST 2 HOUR 15 MIN 8500 ML SIMULATED TREATMENT WAS PERFORMED AND COMPLETED. A DC (DRAIN COMPLICATION ENCOUNTERED) SUPPORT WARNING OCCURRED, AS EXPECTED, WHEN THE PATIENT LINE WAS CLAMPED DURING THE DWELL PHASE CYCLE 1. THE PATIENT SENSOR CALIBRATION CHECK PASSED. THE MUSHROOM HEAD CHECK PASSED. THE CYCLER TESTED POSITIVE FOR GLUCOSE. AN INTERNAL VISUAL INSPECTION OF THE CYCLER FOUND VISUAL EVIDENCE OF DRIED FLUID UNDER PUMP ¿A¿ AND ¿B¿ MUSHROOM HEAD AND WITHIN THE RECESS OF THE BOTTOM COVER ADJACENT TO THE PUMP. THE CAUSE OF THE OBSERVED DRIED FLUID AND FLUID COULD NOT BE DETERMINED. A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NON-CONFORMANCES DURING THE MANUFACTURING PROCESS. IN ADDITION, A DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED AND VERIFIED THAT THE RESULTS OF THE IN-PROGRESS AND FINAL QUALITY CONTROL (QC) TESTING MET ALL REQUIREMENTS. UPON COMPLETION OF THE EVALUATION, THERE WERE NO MALFUNCTIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE CYCLER PERFORMED AS DESIGNED AND AN ASSOCIATED CAUSE COULD NOT BE DETERMINED.
THE PLANT INVESTIGATION IS IN PROCESS. A SUPPLEMENTAL MDR WILL BE SUBMITTED UPON COMPLETION OF THIS ACTIVITY.
IT WAS REPORTED THAT A PERITONEAL DIALYSIS (PD) PATIENT DISCOVERED A FLUID LEAK ON THE INSIDE OF THE CASSETTE DOOR OF THEIR CYCLER DURING THEIR PD TREATMENT ON 11/16/2018. THE PATIENT WAS ADVISED TO DISCONTINUE THE USE OF THEIR CYCLER. UPON FOLLOW UP, THE PERITONEAL DIALYSIS NURSE (PDRN) STATED THE PATIENT IS TRAINED ON PERFORMING STAT DRAINS, AS WELL AS MANUAL PERITONEAL DIALYSIS THERAPY IF NEEDED, AND OPTED NOT TO COMPLETE THEIR TREATMENT ON (B)(6) 2018. THE PATIENT DID NOT DEVELOP ANY SYMPTOMS, ADVERSE EVENTS, INJURIES, OR REQUIRE MEDICAL INTERVENTION AS A RESULT OF THE REPORTED EVENT. THE PATIENT HAS RECEIVED A NEW CYCLER, WHICH IS WORKING WELL AND IS CONTINUING PERITONEAL DIALYSIS THERAPY WITH NO FURTHER ISSUES. THE CASSETTE USED BY THE PATIENT WAS DISCARDED AND IS NOT AVAILABLE FOR RETURN FOR PHYSICAL EVALUATION BY THE MANUFACTURER. THE CYCLER WAS RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 959021 | LIBERTY SELECT CYCLER ASSY(NON-VALUATED) | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | CONCORD MANUFACTURING | 00840861102068 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | DELFLEX PD FLUID| LIBERTY CYCLER SET| DELFLEX PD FLUID| LIBERTY CYCLER SET |