MTS ANTI-IGG CARD
Report
- Report Number
- 1056600-2007-00027
- Event Type
- Death
- Date Received
- February 1, 2007
- Date of Event
- January 10, 2007
- Report Date
- February 1, 2007
- Manufacturer
- MICRO TYPING SYSTEMS, INC.
- Product Code
- KSZ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- UNKNOWN
Narratives
BASED ON THE AVAILABLE INFO PROVIDED BY THE CUSTOMER AND THE RESULTS OF THE INVESTIGATION, THE GEL CARD REACTED AS EXPECTED BY CORRECTLY DETECTING THE PRESENCE OF ANTIBODIES DURING CROSSMATCH TESTING (INCOMPATIBLE). GEL CARD MALFUNCTION COULD NOT BE IDENTIFIED. THE CUSTOMER DID NOT PROVIDE INFO AS TO WHAT INFLUENCED THEIR DECISION TO TRANSFUSE THE PT WITH INCOMPATIBLE BLOOD. THIS COULD NOT BE RULED OUT AS A CONTRIBUTING FACTOR TO THE ADVERSE EVENT. THE CAUSE OF DEATH IS UNKNOWN BY MTS (THIS INFO WAS NOT PROVIDED BY THE CUSTOMER). IF A PT IS TRANSFUSED WITH ANTIGEN-POSITIVE BLOOD, THEY MAY SUFFER HEMOLYTIC TRANSFUSION REACTION. THE LIKELIHOOD OF A SERIOUS INJURY, WHILE NOT FREQUENT, IS NOT REMOTE.
THE CUSTOMER CONTACTED THE CALL CENTER TO INQUIRE ABOUT THE KIND OF REACTIONS OBSERVED WITH ANTI-SDA AND ANTI-HENSHAW. BECAME AWARE OF THIS EVENT THROUGH CONVERSATION WITH THE CUSTOMER. A SAMPLE FROM A PT WITH MULTIPLE ANTIBODIES (ANTI-SDA, ANTI-HENSHAW) REACTED AS MIXED-FIELD DURING COMPATIBILITY/CROSSMATCH TESTING WITH 37 UNITS OF BLOOD. THE UNITS WERE TESTED USING 2 DIFFERENT LOTS OF MTS ANTI-LGG CARDS (LOT #080706001-33 AND 080706001-10). ONE OF THE INCOMPATIBLE UNITS WAS TRANSFUSED AND AFTER RECEIVING APPROXIMATELY 50CC OF BLOOD, THE PT HAD ELEVATED BLOOD PRESSURE AND EXPIRED. THE CUSTOMER DID NOT PROVIDE ANY OTHER INFO REGARDING THE INCIDENT. THE MDR IS BEING FILED TO REPORT THE ADVERSE EVENT REPORTED BY THE CUSTOMER. THERE WAS NO MALFUNCTION OR PRODUCT PROBLEM WITH THE GEL CARD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MTS ANTI-IGG CARD | ANTIBODY SCREENING REAGENT | KSZ | MICRO TYPING SYSTEMS, INC. | NA | 080706001-10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Death |