FDA Adverse Event
Malfunction
Summary report: N
PURITAN BENNETT BB980 VENTILATOR SYSTEM
MDR report key: 8115634
·
Received November 28, 2018
Report
- Report Number
- MW5081711
- Event Type
- Malfunction
- Date Received
- November 28, 2018
- Date of Event
- November 8, 2018
- Report Date
- November 9, 2018
- Manufacturer
- COVIDIEN / MEDTRONIC, FORMERLY NELLCOR PURITAN BENNETT, INC.
- Product Code
- CBK
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- TX, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
COVIDIEN PB 980 VENTILATOR FELL OFF THE STAND WHILE IN TRANSPORT. THE VENTILATOR HAS A LOCKING MECHANISM IN THE BACK THAT HOLDS IT TO THE STAND. OVER TIME, AND WITH THE VIBRATION OF THE UNIT CAUSES THE LOCKING MECHANISM TO SHAKE LOOSE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 950262 | PURITAN BENNETT BB980 VENTILATOR SYSTEM | VENTILATOR CONTINUOUS FACILITY USE | CBK | COVIDIEN / MEDTRONIC, FORMERLY NELLCOR PURITAN BENNETT, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |