FDA Adverse Event Malfunction Summary report: N

AUTO SUTURE PREMIUM MULTIFIRE TA 30-3.5

MDR report key: 81141 · Received April 1, 1997

Report

Report Number
1219930-1997-00752
Event Type
Malfunction
Date Received
April 1, 1997
Date of Event
February 28, 1997
Report Date
March 3, 1997
Manufacturer
UNITED STATES SURGICAL CORP.
Product Code
GAG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE SUBJECT DEVICE REVEALED THAT THE DISTAL END OF THE "#2 LATCHING ROD" LIFTED AND BECAME DISENGAGED FROM THE SUPPORT BLOCK. INTERMITTENTLY, DURING APPROX. OF THE INSTRUMENT, THE LATCH CAM ADVANCED TEH ROD SLIGHTLY FORWARD WHICH RESULTED IN BINDING OF THE ROD BETWEEN THE CAM BRIDGE AND THE SUPPORT BLOCK. THIS INTERMITTENT BINDING PREVENTED THE INSTRUMENT FROM OPENING UPON ACTIVATION OF THE APPROX. BUTTON. AS CORRECTIVE ACTION, PROCESS MODIFICATIONS WERE IMPLEMENTED TO MECHANICALLY PREVENT THE ROD FROM LIFTING AND ELIMINATE FURTHER OCCURRENCE OF THIS CONDITION.

Description of Event or Problem · 1

DURING A LUNG RESECTION PROCEDURE, THE INSTRUMENT JAMMED. HOSP REPORTED THAT NO PT INJURY HAD OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTO SUTURE PREMIUM MULTIFIRE TA 30-3.5 DISPOSABLE SURGICAL STAPLER GAG UNITED STATES SURGICAL CORP. NA N6G69

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN