FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND SPECTRUM

MDR report key: 8112867 · Received November 28, 2018

Report

Report Number
1645337-2018-07091
Event Type
Injury
Date Received
November 28, 2018
Date of Event
October 15, 2018
Report Date
November 8, 2018
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001980
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON 12/27/2018, IT WAS REPORTED TO MENTOR THAT THE DEVICE EVALUATION AND INVESTIGATION HAVE BEEN COMPLETED. UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. WHITE MATERIAL WAS OBSERVED WITHIN THE DEVICE. NO FOREIGN MATERIAL WAS OBSERVED ON THE SHELL SURFACE. MENTOR PRODUCT ANALYSIS LAB DISCOVERED A RENT MEASURING APPROXIMATELY 0.4 CM WITHIN A CREASE AND AN AREA OF SHELL ABRASION ON THE POSTERIOR ASPECT. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE OR THE ETIOLOGY OF THE WHITE MATERIAL FOUND ON THE DEVICE. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BECAUSE MENTOR PRODUCT ANALYSIS LAB WAS UNABLE TO CONFIRM ANY UNUSUAL FAILURE MODE, NO CORRECTIVE ACTION WILL BE TAKEN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 0

ON 11/29/2018, IT WAS REPORTED TO MENTOR THAT THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 0

ON (B)(6) 2018, IT WAS REPORTED TO MENTOR THAT THE DEVICE HISTORY RECORD (DHR) FOR THE LOT NUMBER 233116 HAS BEEN REVIEWED AND IT WAS VERIFIED THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATIONS AND PROCEDURES. (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A DEVICE HISTORY RECORD REVIEW IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: A SALINE MENTOR SMOOTH ROUND SPECTRUM 275CC , CATALOG NUMBER 3501440, LOT NUMBER 233710. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6)-YEAR-OLD FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION PRIMARY WITH SALINE MENTOR SMOOTH ROUND SPECTRUM 275CC BREAST IMPLANTS AND EXPERIENCED DEFLATION ON THE RIGHT BREAST IMPLANT. THE DEFLATION WAS CONFIRMED BY THE HEALTHCARE PROFESSIONAL. AS A RESULT, THE PATIENT HAD UNDERGONE EXPLANTATION ON (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
952377 MENTOR SMOOTH ROUND SPECTRUM PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 233116 00081317001980

Patients

Seq Age Sex Outcome Treatment
1 64 YR Required Intervention