FDA Adverse Event Injury Summary report: N

COCR MODULAR HEAD

MDR report key: 8107654 · Received November 27, 2018

Report

Report Number
0001825034-2018-10876
Event Type
Injury
Date Received
November 27, 2018
Date of Event
October 23, 2017
Report Date
July 29, 2019
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LPH
PMA / PMN Number
K960303
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

REPORTED EVENT WAS CONFIRMED BY REVIEW OF MEDICAL RECORDS. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED. REVISION OP NOTES WERE REVIEWED AND IDENTIFIED PATIENT WAS REVISED DUE TO FAILED CONSTRAINT LINER WITH DISLOCATION OF LEFT TOTAL HIP, FROM DOCUMENTED CHRONIC IMPINGEMENT AND MALALIGNMENT OF THE CUP. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: M2A-MAGNUM PF CUP # ITEM 072830 LOT 072830, UNK CONSTRAINED LINER. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2018-10875, AND 0001825034-2018-10877. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION PROCEDURE APPROXIMATELY SEVEN YEARS POST INITIAL SURGERY DUE TO IMPLANT FRACTURE, DISLOCATION, IMPINGEMENT AND MALPOSITION. THE CUP, LINER AND HEAD WERE REVISED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
945824 COCR MODULAR HEAD PROSTHESIS HIP LPH ZIMMER BIOMET, INC. N/A 196450

Patients

Seq Age Sex Outcome Treatment
1 44 YR Hospitalization| R