FDA Adverse Event Malfunction Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIEC

MDR report key: 810162 · Received October 12, 2006

Report

Report Number
2023826-2006-01506
Event Type
Malfunction
Date Received
October 12, 2006
Report Date
September 14, 2006
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P880091
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

A WORK ORDER SEARCH WAS PERFORMED FOR THE LENS. VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWED THAT ONE LENS HAPTIC IS PARTIALLY TORN OFF AND MISSING. CLEAR SURGICAL AND REDDISH RESIDUE/DEBRIS ON THE PRODUCT. AN EMPTY MTC-60C FP CARTRIDGE TRAY WAS RETURNED WITH LOT # 1182247. A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINT WAS FOUND WITHIN THE SEARCH. CONCLUSION: BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Description of Event or Problem · 1

THE REPORTER STATED THAT WHEN THE AA4203TL SINGLE PIECE SILICONE LENS STERILE POUCH WAS OPENED, IT WAS NOTICED THAT THE LENS HAPTIC WAS BROKEN. NO PT CONTACT OR INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIEC INTRAOCULAR LENS HQL STAAR SURGICAL CO. AA4203TL NA

Patients

Seq Age Sex Outcome Treatment
1 YR CARTRIDGE MODEL MTC-60C, LOT NUMBER 1182247.