FDA Adverse Event Other Summary report: N

ABBOTT TESTPACK RSV

MDR report key: 81016 · Received March 25, 1997

Report

Report Number
1451914-1997-00018
Event Type
Other
Date Received
March 25, 1997
Date of Event
January 20, 1997
Report Date
March 25, 1997
Manufacturer
ABBOTT LABORATORIES
Product Code
GQG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION RESULTS: NO CUSTOMER RETURN RECEIVED. FILE MATERIAL PERFORMED APPROPRIATELY MEETING ALL PERFORMANCE CRITERIA. KNOWN NEGATIVE SAMPLES YIELDED NEGATIVE RESULTS. NEGATIVES WERE CLEAR, CLEAN, AND EASY TO INTERPRET. KNOWN POSITIVES WERE PICKED UP AT APPROPRIATE LEVELS WITH BAR INTENSITIES ACCEPTABLE. IN RESPONSE TO THE CUSTOMER STATING THAT A FAINTLY POSITIVE RESULT IS INTERPRETED BY THE MAJORITY OF TECHS AS NEGATIVE, ANY COLOR ON THE VERTICAL BAR IS A POSITIVE RESULT. SINCE THE CUSTOMER DID NOT RETURN THE KIT, PATIENT SAMPLE, OR A SAMPLE OF THEIR TRANSPORT MEDIA FOR EVALUATION, FURTHER ANALYSIS CANNOT BE PERFORMED. THE COMPLAINT WAS NOT CONFIRMED. EVALUATION COMPLETE-FINAL REPORT.

Description of Event or Problem · 1

THE ACCOUNT STATED THEY OBTAINED DISCREPANT RESULTS WHEN USING THE RSV KIT. THE ACCOUNT STATED THAT WHEN A RAPID ASSAY IS RUN, THE NEGATIVE ASSAYS ARE SENT TO A REFERENCE LAB FOR FA. THE METHOD OF FA IS UNKOWN. THE ACCOUNT STATED THEY HAD A INCIDENT WHICH GAVE A NEGATIVE RESULT ON THE KIT AND POSITIVE ON FA, BUT WHEN THE SAMPLE WAS REPEATED IN THE LAB IT WAS POSITIVE. THE REPORTER ATTRIBUTED THIS TO OPERTOR ERROR AS THE KIT WAS ALLOWED TO SIT OUT AT ROOM TEMPERATURE FOR 5-10 MINUTES BEFORE USE. LABEL STATES KIT REAGENTS AND REACTION DISCS MUST BE AT ROOM TEMPERATURE PRIOR TO USE (MINIMUM 30 MINUTES). ADDITIONAL INFO INDICATES THEY HAD ANOTHER SAMPLE WHICH WAS NEGATIVE ON THE KIT AND POSITIVE WITH FA. ONE PT WAS A BABY ADMITTED WITH UPPER RESPIRATORY PROBLEMS. THE NEGATIVE RESULT FROM RSV KIT IMPACTED PT SINCE TREATMENT WAS DELAYED FOR ABOUT 24 HOURS UNTIL FA RESULT RECEIVED AND PT WAS NOT TREATED OR ISOLATED. NO ADDITIONAL INFO ON PTS AVAILABLE DUE TO CONFIDENTIALITY ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ABBOTT TESTPACK RSV EIA FOR THE RAPID DETECTION OF RSV GQG ABBOTT LABORATORIES NA 24251M100

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other