FDA Adverse Event Injury Summary report: N

VITROS 5,1 FS CHEMISTRY SYSTEM - REFURBISHED

MDR report key: 8101549 · Received November 26, 2018

Report

Report Number
1319681-2018-00118
Event Type
Injury
Date Received
November 26, 2018
Date of Event
October 29, 2018
Report Date
November 26, 2018
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
JJE
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONCLUDED THAT AN OPERATOR SUSTAINED A CUT FINGER WHEN MANIPULATING THE MICROSLIDE INCUBATOR RING ON THE VITROS 5,1 FS CHEMISTRY SYSTEM. THE MOST LIKELY CAUSE OF THE EVENT WAS USER ERROR, AS AN OPERATOR SUSTAINED THE CUT WHEN ATTEMPTING TO CLEAR A SLIDE CARTRIDGE JAM ON THE VITROS 5,1 FS CHEMISTRY SYSTEM. THE OPERATOR SUSTAINED A CUT AS A RESULT OF THIS EVENT, IMMEDIATELY WASHED THEIR HAND AND CLEANED THE CUT WITH ALCOHOL BEFORE APPLYING A BANDAGE. THE OPERATOR WENT TO THE INTERNAL HEALTH CLINIC WHERE GLUE WAS APPLIED TO THE CUT. THE OPERATOR STATED THAT THEY WERE WEARING GLOVES AT THE TIME OF THE EVENT. THERE WAS NO EVIDENCE TO SUGGEST THAT THE VITROS 5,1 FS CHEMISTRY SYSTEM MALFUNCTIONED.

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT AN OPERATOR SUSTAINED A CUT FINGER WHEN MANIPULATING THE MICROSLIDE INCUBATOR RING ON A VITROS 5,1 FS CHEMISTRY SYSTEM. THERE WAS NO ALLEGATION OF SERIOUS OR LONG TERM PERMANENT INJURY AND NO LONG TERM CONSEQUENCES ARE EXPECTED. HOWEVER, THERE IS A POTENTIAL THAT THE OPERATOR MIGHT BE EXPOSED TO BLOOD BORNE PATHOGENS AND SHOULD BE CONSIDERED AS A SERIOUS INJURY. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO).COMPLAINT NUMBERS (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
941972 VITROS 5,1 FS CHEMISTRY SYSTEM - REFURBISHED CHEMISTRY ANALYZER JJE ORTHO-CLINICAL DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1