BD ULTRA-FINE¿ INSULIN SYRINGE
Report
- Report Number
- 1920898-2018-00875
- Event Type
- Malfunction
- Date Received
- November 20, 2018
- Date of Event
- October 31, 2018
- Report Date
- December 4, 2018
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- UDI-DI
- 00382903294169
- PMA / PMN Number
- K170386
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: CUSTOMER RETURNED (12) 1CC, 8MM, 31G SYRINGES (2 LOOSE, 10 IN A SEALED POLY BAG) FROM LOT # 7241532. CUSTOMER STATES THAT THEY ARE HAVING A HARD TIME PRESSING THE PLUNGER AND THERE ARE BURRS ON THE NEEDLES. ALL RETURNED SYRINGES WERE TESTED AND ALL WERE ABLE TO DRAW AND EXPEL PROPERLY WITHOUT ANY OBSERVED DEFECTS. ALL RETURNED SYRINGES WERE ALSO TESTED FOR POINT GEOMETRY, LUBE, AND CANNULA OD (SPECS: OUTER DIAMETER FOR 31G: 0.0100¿-0.0105¿). ALL OBSERVATIONS FALL WITHIN SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 7241532. ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE THREE (3) NOTIFICATIONS [200712892, 200712495, 200712933] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. BASED ON THE SAMPLES / PHOTO(S) RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED.
IT WAS REPORTED THAT DURING USE OF THE BD ULTRA-FINE¿ INSULIN SYRINGE THE PLUNGER ROD IS HARD TO MOVE.
INVESTIGATION SUMMARY : NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 7241532. ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. INVESTIGATION CONCLUSION : BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED. ROOT CAUSE DESCRIPTION : ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED AS NO SAMPLES OR PHOTOS WERE RETURNED. RATIONALE : BASED ON THE INVESTIGATION, NO ADDITIONAL INVESTIGATION AND NO CAPA IS REQUIRED AT THIS TIME.
IT WAS REPORTED THAT DURING USE OF THE BD ULTRA-FINE¿ INSULIN SYRINGE THE PLUNGER ROD IS HARD TO MOVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 930932 | BD ULTRA-FINE¿ INSULIN SYRINGE | INSULIN SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 7241532 | 00382903294169 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |