FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 8086766 · Received November 19, 2018

Report

Report Number
2024168-2018-08932
Event Type
Injury
Date Received
November 19, 2018
Date of Event
October 31, 2018
Report Date
December 4, 2018
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION. THE OTHER PROGLIDE DEVICE IS FILED UNDER A SEPARATE MEDWATCH REPORT NUMBER.

Additional Manufacturer Narrative · 1

(B)(4). LOT NUMBER UPDATED FROM 8071741 TO 7100942. VISUAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED CUFF MISS COULD NOT BE TESTED DUE TO DEVICE COMPONENTS NOT BEING RETURNED. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A CALCIFIED COMMON FEMORAL ARTERY WAS ATTEMPTED USING TWO PROGLIDE DEVICES VIA A 16FR SHEATH, AFTER A VASCULAR SURGEON INTERVENTIONAL PROCEDURE. REPORTEDLY, WHEN REMOVING THE PLUNGER, THE SUTURE COULD NOT BE FOUND WITH BOTH PROGLIDE DEVICES. MANUAL ARTERIAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
926214 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 7100942

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention