G7 NEUTRAL ARCOMXL LNR 36MM H
Report
- Report Number
- 0001825034-2018-10779
- Event Type
- Injury
- Date Received
- November 19, 2018
- Date of Event
- October 23, 2018
- Report Date
- February 5, 2019
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- PBI
- PMA / PMN Number
- K121874
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. VISUAL INSPECTION FOUND THE RIM TO BE GOUGED AND A SCREW HOLE PRESENT. THE SCREW HOLE EXTENDS INTO THE OUTER RADIUS. PITTING AND SCRATCHING WAS ALSO OBSERVED ON THE OUTER RADIUS. THE BARB IS SCRAPED SUCH THAT A STRAND OF POLY PROTRUDES FROM THE LINER. NO DIMENSIONAL ANALYSIS WILL BE PERFORMED DUE TO THE IMPLANTATION AND REMOVAL OF THE LINER. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
(B)(4). PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: ITEM # 010000668, G7 PPS LTD ACET SHELL, LOT # 6081830 ITEM # 010000743, G7 NEUTRAL ARCOMXL LNR , LOT # 3517308, ITEM # 010000743, G7 NEUTRAL ARCOMXL LNR , LOT # 3575886. MULTIPLE REPORTS HAVE BEEN SUBMITTED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORTS: 0001825034-2018-10777, 0001825034-2018-10778, 0001825034-2018-10779, 0001825034-2018-10780.
IT WAS REPORTED THAT MULTIPLE LINERS WOULD NOT SEAT IN CUP. AFTER THE THIRD LINER, THE SURGEON REMOVED THE CUP, AND IMPLANTED AN ALTERNATE CUP. THE THIRD LINER WAS THEN USED AND SEATED SUCCESSFULLY. THE MALFUNCTION CAUSED A 45 MINUTE SURGICAL DELAY. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 923734 | G7 NEUTRAL ARCOMXL LNR 36MM H | HIP PROSTHESIS | PBI | ZIMMER BIOMET, INC. | N/A | 3822583 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |