BD POSIFLUSH¿ NS FILLED SYRINGE
Report
- Report Number
- 9616657-2018-00053
- Event Type
- Malfunction
- Date Received
- November 19, 2018
- Date of Event
- October 31, 2018
- Report Date
- December 5, 2018
- Manufacturer
- BECTON, DICKINSON AND CO.
- Product Code
- NGT
- UDI-DI
- 30382903065463
- PMA / PMN Number
- K141311
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHARMACIST
Narratives
THIRTEEN SAMPLE UNITS BELONGING TO LOT NUMBER 7110587 WERE RECEIVED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. UPON VISUAL INSPECTION OF THE SAMPLES, PLUNGER RESISTANCE WAS IDENTIFIED. A DEVICE HISTORY RECORD REVIEW FOR LOT NUMBER 7110587 REVEALED ONE NON-CONFORMANCE DURING THE PRODUCTION PROCESS THAT MAY HAVE CONTRIBUTED TO THIS INCIDENT. THERE WAS A RECORDED INTERMITTENT ISSUE WITH THE HOSING IN THE MANUFACTURING MACHINERY AT THE TIME OF THIS LOT'S PRODUCTION. PRODUCT ASSOCIATED WITH THIS DEFECT WAS HELD FOR INSPECTION AND ALL AFFECTED MATERIAL SHOULD HAVE BEEN REJECTED TO SCRAP. IT IS POSSIBLE THAT THE PRODUCT REPORTED WITHIN THIS INCIDENT WENT UNDETECTED. A CORRECTION HAS BEEN MADE TO THE SILICONE SUPPLY HOSING MACHINERY WITHIN THE DESIGNATED FILL ROOM TO PREVENT THIS ISSUE FROM REOCCURRING. COMPLAINTS RECEIVED FOR THIS DEVICE AND DEFECT WILL BE MONITORED BY OUR QUALITY TEAM FOR EMERGING TRENDS.
IT WAS REPORTED THAT THE BD POSIFLUSH NS FILLED SYRINGES WERE "ONLY ALLOWING ABOUT 5ML OF THE 10ML TO BE DISPENSED" DUE TO DIFFICULT PLUNGER MOVEMENT. THERE WAS NO REPORT OF EXPOSURE, INJURY, OR MEDICAL INTERVENTION.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD POSIFLUSH¿ NS FILLED SYRINGES WERE "ONLY ALLOWING ABOUT 5ML OF THE 10ML TO BE DISPENSED" DUE TO DIFFICULT PLUNGER MOVEMENT. THERE WAS NO REPORT OF EXPOSURE, INJURY, OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 924250 | BD POSIFLUSH¿ NS FILLED SYRINGE | INTRAVASCULAR CATHETER | NGT | BECTON, DICKINSON AND CO. | 7110587 | 30382903065463 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |