FDA Adverse Event Injury Summary report: N

CYSTO-NEPHRO VIDEOSCOPE

MDR report key: 8084684 · Received November 19, 2018

Report

Report Number
2951238-2018-00708
Event Type
Injury
Date Received
November 19, 2018
Report Date
March 1, 2019
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FAJ
UDI-DI
04953170310522
PMA / PMN Number
K062049
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO MAKE A CORRECTION ON THE PROCODE FROM NWB TO FAJ.

Additional Manufacturer Narrative · 1

THE SCOPE HAS NOT RETURNED TO OLYMPUS FOR EVALUATION. THE REPORTED COMPLAINT CANNOT BE CONFIRMED. DESPITE SEVERAL FOLLOWUP ATTEMPTS, NO FURTHER INFORMATION IS CURRENTLY AVAILABLE REGARDING THE INFECTION, THE PROCEDURE INVOLVED, THE PATIENT OUTCOME, THE SPECIFIC SCOPE INVOLVED, ITS PRE-PROCEDURE INSPECTION, OR WHETHER THE SCOPE DEVICE IS RETURNING TO OLYMPUS FOR EVALUATION. AN OLYMPUS ENDOSCOPY SUPPORT SPECIALIST HAS SINCE PROVIDED RETRAINING TO ADDRESS THE REPROCESSING ISSUES THAT WERE A POTENTIAL CONTRIBUTING FACTOR. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED ACCORDINGLY.

Description of Event or Problem · 1

1 OF 3. DURING A CUSTOMER EDUCATION SESSION ON MANUAL CLEANING AND DISINFECTION, OLYMPUS WAS INFORMED THAT THE USER FACILITY HAD EXPERIENCED A PRIOR INFECTION INCIDENT INVOLVING THIS MODEL SCOPE. THE ACTUAL SCOPE SERIAL NUMBER WAS NOT DISCLOSED. THE INFECTION TYPE, DATE OF PROCEDURE AND PATIENT DETAILS WERE NOT DISCLOSED. IT WAS REPORTED THAT PRIOR TO THE EDUCATION SESSION, THE SCOPE REPROCESSING AT THE USER FACILITY HAD NOT BEEN PERFORMING THE LEAK TEST, WAS NOT COMPLETING PRE-CLEANING, DID NOT HAVE THE APPROPRIATE CLEANING BRUSHES FOR THE ENDOSCOPE CHANNELS, DID NOT TAKE THE IRRIGATION PORT OFF BETWEEN USES, AND DID NOT COMPLETELY IMMERSE THE ENDOSCOPE IN HIGH LEVEL DISINFECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
924733 CYSTO-NEPHRO VIDEOSCOPE CYSTO-NEPHRO VIDEOSCOPE FAJ OLYMPUS MEDICAL SYSTEMS CORP. CYF-VHR 04953170310522

Patients

Seq Age Sex Outcome Treatment
1 Other