FDA Adverse Event
Malfunction
Summary report: N
FFRCT
MDR report key: 8083832
·
Received November 19, 2018
Report
- Report Number
- 3011276938-2018-00008
- Event Type
- Malfunction
- Date Received
- November 19, 2018
- Date of Event
- October 9, 2018
- Report Date
- November 19, 2018
- Manufacturer
- HEARTFLOW, INC.
- Product Code
- PJA
- UDI-DI
- 00853341006015
- PMA / PMN Number
- K161772
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 501
Narratives
Additional Manufacturer Narrative · 1
AS PART OF HEARTFLOW'S QUALITY MONITORING PROCESS, WE IDENTIFIED A POTENTIAL FALSE NEGATIVE AND INFORMED THE ORDERING PHYSICIAN. HFID # (B)(4): THE INVESTIGATION IDENTIFIED A POTENTIAL FALSE NEGATIVE IN THE RCA. THIS WAS DUE TO ANALYST ERROR; WHEN EDITED THE REASSESSMENT OF THE ANALYSIS SHOWED A DECREASE IN THE DISTAL FFRCT VALUE FROM 0.81 TO 0.68. WHILE HEARTFLOW HAS IDENTIFIED THIS ISSUE, THE PHYSICIAN HAS NOT PROVIDED FEEDBACK, NOR INDICATED A SAFETY EVENT WITH THE PATIENT. NO ADDITIONAL FOLLOW-UP TO THIS MDR IS EXPECTED.
Description of Event or Problem · 1
HEARTFLOW IDENTIFIED A POTENTIAL FALSE NEGATIVE RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 924951 | FFRCT | HEARTFLOW FFRCT | PJA | HEARTFLOW, INC. | FFRCT 2.24 | 00853341006015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |