FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD-PAK

MDR report key: 8083627 · Received November 19, 2018

Report

Report Number
3004123209-2018-00704
Event Type
Malfunction
Date Received
November 19, 2018
Date of Event
October 1, 2018
Report Date
December 6, 2018
Manufacturer
HEARTSINE TECHNOLOGIES LTD
Product Code
MKJ
PMA / PMN Number
K014067
Removal / Correction Number
Z-0124-2013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HISTORY RECORDS FOR THE SAM 300P DEVICE WAS REVIEWED AND THIS CONFIRMED THAT ALL MANUFACTURING AND QUALITY CHECKS AND TESTS HAD BEEN SUCCESSFULLY COMPLETED. NO REWORK WAS CONDUCTED. THE SAM 300P PASSED ¿OUT QAT FROM HEARTSINE TECHNOLOGIES ON THE (B)(6) 2010. THE RANDOM NATURE OF THE EVENTS STORED IN THE MEMORY AND THE 10 MINUTE TIMEOUTS, WOULD SUGGEST THE DEVICE WAS SWITCHING ON AUTOMATICALLY. EXPERIENCE HAS SHOWN THAT IT IS REASONABLE TO CONCLUDE THAT THESE SYMPTOMS MAY BE ATTRIBUTED TO A MEMBRANE FAILURE. THE RETURNED SAM 300P IS NOW OUTSIDE OF ITS WARRANTY PERIOD AND WILL BE SCRAPPED.

Description of Event or Problem · 0

MEMBRANE FAILURE DETECTED. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY RECORDS FOR THE SAM 300P DEVICE WAS REVIEWED AND THIS CONFIRMED THAT ALL MANUFACTURING AND QUALITY CHECKS AND TESTS HAD BEEN SUCCESSFULLY COMPLETED. THE SAM 300P PASSED ¿OUT QAT FROM HEARTSINE TECHNOLOGIES ON THE (B)(6) 2010. THE RANDOM NATURE OF THE EVENTS STORED IN THE MEMORY AND THE 10 MINUTE TIMEOUTS, WOULD SUGGEST THE DEVICE WAS SWITCHING ON AUTOMATICALLY. EXPERIENCE HAS SHOWN THAT IT IS REASONABLE TO CONCLUDE THAT THESE SYMPTOMS MAY BE ATTRIBUTED TO A MEMBRANE FAILURE. THE RETURNED SAM 300P IS NOW OUTSIDE OF ITS WARRANTY PERIOD AND WILL BE SCRAPPED.

Description of Event or Problem · 1

MEMBRANE FAILURE DETECTED. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
923832 HEARTSINE SAMARITAN 300P AND PAD-PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES LTD

Patients

Seq Age Sex Outcome Treatment
1