AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE
Report
- Report Number
- 2183959-2018-00128
- Event Type
- Injury
- Date Received
- November 16, 2018
- Date of Event
- November 2, 2018
- Report Date
- September 15, 2016
- Manufacturer
- BOSTON SCIENTIFIC - MINNETONKA
- Product Code
- FHW
- UDI-DI
- 00878953003610
- PMA / PMN Number
- N970012
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
MODEL NUMBER: 72404253, SERIAL NUMBER: (NULL), LOT NUMBER: 103752005, MODEL DESCRIPTION: LGX PRECONNECT MS 21CM PS, IZ, IMPLANT DATE: (B)(6) 2015, EXPLANT DATE: (B)(6) 2018, EXPIRATION DATE: 09/22/2017, MANUFACTURE DATE: 10/08/2015, (B)(4). MODEL NUMBER: 72404134, SERIAL NUMBER: (NULL), LOT NUMBER: 923910008, MODEL DESCRIPTION: BELT CUFF FGS, 6.0 CM, IZ, IMPLANT DATE: (B)(6) 2015. EXPLANT DATE: (B)(6) 2018, EXPIRATION DATE: 02/24/2016, MANUFACTURE DATE: 03/10/2015, (B)(4).
IT WAS REPORTED THAT THE PATIENT'S INFLATABLE PENILE PROSTHESIS PUMP COMPONENT COLLAPSED AND WOULD NOT ACTIVATE AFTER USING THE DEFLATE BUTTON. THE PATIENT SAW THEIR PHYSICIAN AND REPORTED THE COLLAPSED PUMP AND THAT THE PROBLEM HAS PROGRESSIVELY GOTTEN WORSE OVER THE LAST FEW WEEKS. THE PATIENT TRIED TO RE-CYCLE THE PUMP USING THE DEFLATE BUTTON AND HAD SOME SUCCESS BUT AGAIN EXPERIENCED ISSUES ACTIVATING. NO ADDITIONAL PATIENT COMPLICATIONS WERE REPORTED IN RELATION TO THIS EVENT. IT WAS SUBSEQUENTLY REPORTED THE IMPLANT WAS REMOVED AN REPLACED DUE TO PUMP FAILURE AND STICKY PUMP. THIS COMPLAINT WAS INITIALLY SUBMITTED TO FDA VIA ASR REPORT Q3 2016 AND ADDITIONAL INFORMATION WAS RECEIVED BY BOSTON SCIENTIFIC. DUE TO THE REMOVAL OF EXEMPTION E1997037, THIS INFORMATION IS PROVIDED VIA SUPPLEMENTAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 917851 | AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE | DEVICE, IMPOTENCE, MECHANICAL/HYDRAULIC | FHW | BOSTON SCIENTIFIC - MINNETONKA | 72404253 | 00878953003610 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Hospitalization| R |