FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 8080394 · Received November 16, 2018

Report

Report Number
2649622-2018-18868
Event Type
Injury
Date Received
November 16, 2018
Date of Event
October 26, 2018
Report Date
November 16, 2018
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS; HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE OF THE SUPERIOR VENA CAVA DEFIBRILLATION COIL WAS BEYOND THE EXPECTED UPPER RANGE. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE TREND OF THE SUPERIOR VENA CAVA DEFIBRILLATION COIL WAS VARIABLE. CONCOMITANT MEDICAL PRODUCTS: 5076-52 LEAD, IMPLANTED: (B)(6) 2011. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH SUPERIOR VENA CAVA (SVC) IMPEDANCE. THE LEAD WAS PROGRAMMED OFF AND REMAINS IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
920536 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention DDMC3D4 ICD