Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

VERAX REQUESTED: SAMPLES FROM THE PLATELET UNITS. THE USER FACILITY HAS NOT PROVIDED SAMPLES. SLANTS CONTAINING THE MICROORGANISMS IDENTIFIED FROM THE PLATELET UNITS. THE HOSPITAL PROVIDED SLANTS. PHOTOGRAPHS OR THE ABILITY TO VIEW THE PGD TEST DEVICES THAT WERE USED TO TEST THE PLATELET UNIT, THE RESIDUAL SAMPLES, ETC. THE USER FACILITY HAS NOT PROVIDED ANY PHOTOS OR ALLOWED VIEWING OF ANY THE COMPLETED TEST DEVICES. VERAX REQUESTED: PHOTOGRAPHS OF RESULTS OF CULTURE PLATES VERAX HAS NOT HAD ACCESS TO THIS INFORMATION. ADDITIONAL ACCESS TO LABORATORY PERSONNEL SO THAT HUMAN-TO-HUMAN TROUBLESHOOTING COULD BE PERFORMED. VERAX HAS NOT BEEN ABLE TO SCHEDULE SUCH ACTIVITIES. VERAX PERFORMED AN INVESTIGATION USING REPRESENTATIVE PLATELET PGD TEST DEVICES, THE ACINETOBACTER BAUMANNII STRAIN(S) PROVIDED BY THE HOSPITAL AND IN-DATE PLATELETS PURCHASED FOR IN VITRO DIAGNOSTIC USE. A. BAUMANNII WAS GROWN ON BLOOD AGAR PLATES. LIVE BACTERIA WERE THEN SPIKED INTO PLATELETS AND TESTED WITH THE PLATELET PGD TEST. THE PGD TEST WAS ABLE TO DETECT THE ORGANISM IDENTIFIED AT THE USER FACILITY. AS THE USER FACILITY WAS NOT ABLE TO DETECT THE ORGANISM USING THE PGD TEST, VERAX CONCLUDES THAT THE CONCENTRATIONS OF BACTERIA PRESENT IN THE PLATELET UNITS TESTED AT THE HOSPITAL WERE BELOW THE LEVEL AT WHICH VERAX DETECTED THE ORGANISM. VERAX DOES NOT KNOW THE CONCENTRATION OF BACTERIA IN THE SAMPLES TESTED AT THE USER FACILITY. NOTE: THE USER FACILITY REPORT INCLUDED A "SERIAL NUMBER" ENTRY. THE PLATELET PGD TEST DOES NOT HAVE SERIAL NUMBERS. THE VERAX PLATELET PGD TEST INSTRUCTIONS FOR USE SECTION M, LIMITATIONS, 9, STATES THAT "A NON-REACTIVE TEST RESULT DOES NOT MEAN THE UNIT IS STERILE OR BACTERIA FREE. NON-REACTIVE RESULTS MAY OCCUR IF..." SECTION M.9.C..PROVIDES ONE EXAMPLE AS FOLLOWS: "THE CONCENTRATION OF BACTERIA IS BELOW THE LIMIT OF DETECTION OF THE TEST".

Description of Event or Problem · 1

ON OCTOBER 1, THE BLOOD CENTER MANUFACTURED TWO (2) APHERESIS PLATELET UNITS IN PLATELET ADDITIVE SOLUTION/PLASMA FROM A SINGLE COLLECTION. PRIOR TO DISTRIBUTION, PLATELETS ARE SUBJECTED TO PRIMARY TESTING FOR BACTERIA USING A GROWTH BASED QUALITY CONTROL TEST AND UNITS ARE DISTRIBUTED AS 'NEGATIVE TO DATE.' THE HOSPITAL RECEIVED BOTH UNITS THE EVENING OF OCTOBER 3. ON THE MORNING OF OCTOBER 4, HOSPITAL USED THE PLATELET PGD TEST AS A SECONDARY TEST TO DETECT BACTERIA. THE PGD RESULTS WERE NON-REACTIVE. THE UNITS REPORTEDLY LOOKED "NORMAL." THE UNITS WAS TRANSFUSED TO HEMATOPOIETIC STEM CELL TRANSPLANT RECIPIENTS. THE RECIPIENTS EXPERIENCED SERIOUS SEPTIC TRANSFUSION REACTIONS AFFECTING THEIR KIDNEYS. THE UNITS WERE RETURNED TO THE LAB WHERE SAMPLES WERE WITHDRAWN AND GRAM STAINED. BLOOD CULTURES WERE ALSO PERFORMED. GRAM STAINS WERE POSITIVE. BLOOD CULTURES WERE POSITIVE IN ABOUT 2-3 HOURS. ACINETOBACTER BAUMANNII, A GRAM-NEGATIVE ORGANISM, WAS IDENTIFIED. BLOOD SAMPLES FROM THE RECIPIENTS WERE TESTED AND DETERMINED TO CONTAIN THE SAME BACTERIA AS IN THE PLATELET UNITS. DISCARDED PLATELET SAMPLES WERE RETRIEVED AND TESTED WITH THE PLATELET PGD TEST. THE PGD RESULTS WERE NON-REACTIVE. DURING A PHONE CONVERSATION ON OCT. 10, A FACILITY REPRESENTATIVE REPORTED THAT THE PATIENTS HAD IMPROVED. NO FURTHER INFORMATION HAS BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
921506 PLATELET PGD TEST BACTERIAL DETECTION SYSTEM FOR PLATELETS FOR TRANSFUSION MZC VERAX BIOMEDICAL INCOPORATED 17355-001

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening