FDA Adverse Event Malfunction Summary report: N

D101 KIDS INFANT HOLLOW FIBER MEMBRANE OXYGENATOR W/ INTEGRATED RESERVOIR

MDR report key: 8072849 · Received November 15, 2018

Report

Report Number
9680841-2018-00030
Event Type
Malfunction
Date Received
November 15, 2018
Date of Event
October 10, 2018
Report Date
February 14, 2019
Manufacturer
SORIN GROUP ITALIA S.R.L.
Product Code
DTZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THERE WAS NO PATIENT INVOLVEMENT. THE COMPLAINED D101 KIDS HOLLOW FIBER OXYGENATOR (CATALOG NUMBER 050540J, LOT 1608250105) IS NOT DISTRIBUTED IN THE USA, THEREFORE THE UDI IS NOT APPLICABLE. THE OXYGENATOR ITEM 050540J IS SIMILAR TO THE D101 KIDS OXYGENATOR 050540, WHICH IS DISTRIBUTED IN THE USA, FOR WHICH THE DEVICE IDENTIFIER IS (B)(4). THE AGE OF THE DEVICE WAS CALCULATED AS THE TIME ELAPSED BETWEEN DEVICE STERILIZATION AND THE DATE OF THE EVENT. (B)(4). THE PRODUCT ITEM 050540J IS NOT DISTRIBUTED IN THE USA, BUT IT IS SIMILAR TO THE D101 KIDS OXYGENATOR 050540, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K072091). THE DEVICE MANUFACTURE DATE REFERS TO MANUFACTURE DATE OF THE STERILE, FINISHED OXYGENATOR. SORIN GROUP (B)(4) MANUFACTURES THE D101 KIDS HOLLOW FIBER OXYGENATOR. THE INCIDENT OCCURRED IN (B)(6). PER EXEMPTION NUMBER E2016005. THE INVOLVED RESERVOIR HAS BEEN REQUESTED FOR RETURN TO SORIN GROUP (B)(4) FOR INVESTIGATION. THE RESERVOIR WAS ASSEMBLED INTO A CUSTOMIZED PERFUSION TUBING CIRCUIT BY A NON-SORIN ASSEMBLER IN (B)(4). IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT. DEVICE NOT YET RETURNED.

Additional Manufacturer Narrative · 0

SORIN GROUP ITALIA MANUFACTURES THE D101 KIDS HOLLOW FIBER OXYGENATOR. THE INCIDENT OCCURRED IN FUCHU, JAPAN. PER EXEMPTION NUMBER E2016005, SORIN GROUP ITALIA S.R.L. IS SUBMITTING THE REPORT FOR BOTH SORIN GROUP ITALIA S.R.L (MANUFACTURER) AND SORIN GROUP USA., INC. (IMPORTER). INSPECTION OF THE RETURNED RESERVOIR CONFIRMED THE BLOOD OUT CONNECTOR COMPLETELY SEPARATED IN TWO PARTS. NO MANUFACTURING DEFECT NOR DIMENSION DEFECT COULD BE IDENTIFIED IN THE RETURNED RESERVOIR. THE TUBE ASSEMBLED TO THE RESERVOIR WAS A NON-SORIN TUBE ASSEMBLED BY A NON-SORIN ASSEMBLER IN JAPAN. A REVIEW OF THE DHR DID NOT IDENTIFY ANY DEVIATIONS OR NON-CONFORMITIES RELEVANT TO THE REPORTED ISSUE. AS NO DEVICE ISSUE COULD BE IDENTIFIED AND THIS IS THE UNIQUE CASE IN THE LAST 12 MONTHS, NO CORRECTIVE ACTION IS DEEMED NECESSARY. LIVANOVA WILL KEEP MONITORING THE MARKET FOR SIMILAR EVENTS.

Description of Event or Problem · 0

SORIN GROUP (B)(4) RECEIVED A REPORT THAT THE RESERVOIR BLOOD-OUTLET CONNECTOR OF THE D101 KIDS HOLLOW FIBER OXYGENATOR WAS HIT AND IT WAS COMPLETELY BROKEN OFF DURING DEBUBBLING OF THE CIRCUIT. THE ISSUE OCCURRED BEFORE USE DURING SET-UP AND PRIMING.