FDA Adverse Event Malfunction Summary report: N

CDS/R/4K DILUENT/SHEATH, 20L

MDR report key: 807049 · Received January 8, 2007

Report

Report Number
2919069-2007-00011
Event Type
Malfunction
Date Received
January 8, 2007
Date of Event
October 7, 2006
Report Date
December 15, 2006
Manufacturer
ABBOTT DIAGNOSTICS DIV/CELLDYN
Product Code
GKL
PMA / PMN Number
k961439
Removal / Correction Number
2919069-1/4/07-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

USE OF DILUENT SHEATH LOT 42063I2 CAN CAUSE ERRORS IN RESULTS WHEN USED ON THE CELL-DYN RUBY. CELL-DYN SAPPHIRE AND CELL-DYN 4000. RESULTS THAT MAY BE IMPACTED INCLUDE RBC, MCV, RDW, PLATELETS AND MPV. THE IMPACT TO RESULTS VARIES BY INSTRUMENT TYPE AS WELL AS THE MATERIAL TESTED, I.E., CONTROL, CONTROL TYPE OR WHOLE BLOOD. RESULTS THAT MAY BE IMPACTED INCLUDE SMALL (LESS THAN 5%) SHIFTS IN RBC, MCV, RDW, AND MPV ALONG WITH 10-20% SHIFT IN OPTICAL PLATELET COUNTS. THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE ACCOUNT STATED THAT MCV, MCHC AND PLATELET OPTICAL IS OUT OF RANGE ON MULTIPLE LEVELS OF QC WHEN TESTED ON THE CELL-DYN 4000 ANALYZER. THERE WAS NO REPORT OF INCORRECT PT RESULTS OR IMPACT TO PT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CDS/R/4K DILUENT/SHEATH, 20L DILUENT/SHEATH FOR USE ON THE CELL-DYN SAPPHIRE, CELL-DYN 40 GKL ABBOTT DIAGNOSTICS DIV/CELLDYN NA 42063I2

Patients

Seq Age Sex Outcome Treatment
1 NA CELL-DYN 4000 ANALYZER LIST #1H01-01