CDS/R/4K DILUENT/SHEATH, 20L
Report
- Report Number
- 2919069-2007-00011
- Event Type
- Malfunction
- Date Received
- January 8, 2007
- Date of Event
- October 7, 2006
- Report Date
- December 15, 2006
- Manufacturer
- ABBOTT DIAGNOSTICS DIV/CELLDYN
- Product Code
- GKL
- PMA / PMN Number
- k961439
- Removal / Correction Number
- 2919069-1/4/07-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
USE OF DILUENT SHEATH LOT 42063I2 CAN CAUSE ERRORS IN RESULTS WHEN USED ON THE CELL-DYN RUBY. CELL-DYN SAPPHIRE AND CELL-DYN 4000. RESULTS THAT MAY BE IMPACTED INCLUDE RBC, MCV, RDW, PLATELETS AND MPV. THE IMPACT TO RESULTS VARIES BY INSTRUMENT TYPE AS WELL AS THE MATERIAL TESTED, I.E., CONTROL, CONTROL TYPE OR WHOLE BLOOD. RESULTS THAT MAY BE IMPACTED INCLUDE SMALL (LESS THAN 5%) SHIFTS IN RBC, MCV, RDW, AND MPV ALONG WITH 10-20% SHIFT IN OPTICAL PLATELET COUNTS. THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE ACCOUNT STATED THAT MCV, MCHC AND PLATELET OPTICAL IS OUT OF RANGE ON MULTIPLE LEVELS OF QC WHEN TESTED ON THE CELL-DYN 4000 ANALYZER. THERE WAS NO REPORT OF INCORRECT PT RESULTS OR IMPACT TO PT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CDS/R/4K DILUENT/SHEATH, 20L | DILUENT/SHEATH FOR USE ON THE CELL-DYN SAPPHIRE, CELL-DYN 40 | GKL | ABBOTT DIAGNOSTICS DIV/CELLDYN | NA | 42063I2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | CELL-DYN 4000 ANALYZER LIST #1H01-01 |