HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY
Report
- Report Number
- 3007042319-2018-05543
- Event Type
- Malfunction
- Date Received
- November 14, 2018
- Date of Event
- November 2, 2018
- Report Date
- March 5, 2019
- Manufacturer
- HEARTWARE, INC.
- Product Code
- DSQ
- UDI-DI
- 00888707000369
- PMA / PMN Number
- P100047
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE BATTERY WAS RETURNED FOR EVALUATION. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICES IN RELATION TO THE REPORTED EVENT. FAILURE ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE BATTERY PASSED VISUAL EXAMINATION. FUNCTIONAL TESTING REVEALED THAT THE BATTERY WAS UNABLE TO CHARGE DUE TO A FLAG. FURTHER ANALYSIS REVEALED THAT THE FLAG WAS DUE TO AN OVER-VOLTAGE FAULT BY THE ANALOG FRONT END (AFE) THAT WAS TRIGGERED WHEN A CELL PAIR INCREASED ABOVE THE VOLTAGE THRESHOLD. WHEN AN OVER-VOLTAGE FAULT OCCURS, THE BATTERY IS UNABLE TO CHARGE BUT CAN STILL PROVIDE POWER. THE FLAG WAS REMOVED AFTER ALLOWING THE BATTERY TO DISCHARGE, CAUSING THE VOLTAGE TO DECREASE BELOW THE VOLTAGE THRESHOLD. THE BATTERY DISCHARGED AS EXPECTED DURING FUNCTIONAL TESTING. LOG FILE ANALYSIS ALSO REVEALED THAT THE BATTERY DISCHARGED AS EXPECTED WHEN IN USE. THE MOST LIKELY ROOT CAUSE CAN BE ATTRIBUTED TO AN OVERVOLTAGE FAULT BY THE ANALOG FRONT END (AFE) INTEGRATED CIRCUIT. THE REPORTED "POWER CONSUMPTION" ISSUE COULD NOT BE CONFIRMED DURING BENCH TESTING. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. CONCOMITANT MEDICAL PRODUCTS: W1TR01 ICD, THREE 4968-35 LEADS, IMPLANTED: (B)(6) 2018. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE BATTERY WAS NOT HOLDING ITS CHARGE AND DRAINING TO A RED SIGNAL IN TWO TO THREE HOURS. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 910098 | HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY | VENTRICULAR (ASSISST) BYPASS | DSQ | HEARTWARE, INC. | 1650DE | 00888707000369 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR | 1103 VAD |