CDS/R/4K DILUENT/SHEATH, 20L
Report
- Report Number
- 2919069-2007-00008
- Event Type
- Malfunction
- Date Received
- January 4, 2007
- Date of Event
- October 18, 2006
- Report Date
- December 15, 2006
- Manufacturer
- ABBOTT DIAGNOSTICS DIV/CELLDYN
- Product Code
- GKL
- PMA / PMN Number
- k961439
- Removal / Correction Number
- 2919069-1/4/07-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
USE OF DILUENT SHEATH LOT 42063I2 CAN CAUSE ERRORS IN RESULTS WHEN USED ON THE CELL-DYN RUBY. CELL-DYN SAPPHIRE AND CELL-DYN 4000. RESULTS THAT MAY BE IMPACTED INCLUDE RBC, MCV, RDW, PLATELETS AND MPV. THE IMPACT TO RESULTS VARIES BY INSTRUMENT TYPE AS WELL AS THE MATERIAL TESTED, I.E., CONTROL, CONTROL TYPE OR WHOLE BLOOD. RESULTS THAT MAY BE IMPACTED INCLUDED SMALL (LESS THAN 5%) SHIFTS IN RBC, MCV, RDW, AND MPV ALONG WITH 10-20% SHIFT IN OPTICAL PLATELET COUNTS. THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE CUSTOMER STATES THAT CONTROLS WERE OUT OF RANGE HIGH AND LOW FOR MULTIPLE LEVELS. THE CUSTOMER STATES THAT THE ISSUE STARTED WHEN A NEW DILUENT/SHEATH REAGENT LOT #42063I2 WAS INSTALLED ON THE CELL-DYN 4000 ANALYZER. THERE WERE NO PT RESULTS AFFECTED OR REPORTED OUT OF THE LAB. NO IMPACT TO PT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CDS/R/4K DILUENT/SHEATH, 20L | DILUENT/SHEATH FOR USE ON THE CELL-DYN SAPPHIRE, CELL-DYN 40 | GKL | ABBOTT DIAGNOSTICS DIV/CELLDYN | NA | 42063I2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO | CELL-DYN 4000 ANALYZER LIST NO.1H01-01 |