FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES

MDR report key: 8067896 · Received November 13, 2018

Report

Report Number
1024879-2018-01373
Event Type
Malfunction
Date Received
November 13, 2018
Date of Event
October 19, 2018
Report Date
January 4, 2019
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903679862
PMA / PMN Number
BK050036
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD RECEIVED SAMPLES AND PHOTOS FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER¿S INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, EVALUATION/TESTING OF THE CUSTOMER SAMPLES WAS PERFORMED AND POOR BARRIER SEPARATION WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH A CAPA AND POTENTIAL CAUSES HAVE BEEN IDENTIFIED. AS A RESULT, CORRECTIVE ACTIONS HAVE BEEN ESTABLISHED AND ARE IN THE PROCESS OF BEING IMPLEMENTED. INVESTIGATION CONCLUSION: BASED ON EVALUATION OF THE CUSTOMER PHOTOS, THE CUSTOMER¿S INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, EVALUATION/TESTING OF THE CUSTOMER SAMPLES WAS CONDUCTED AND POOR BARRIER SEPARATION WAS OBSERVED. FURTHER INVESTIGATION ACTIVITIES HAVE BEEN CONDUCTED THROUGH A CAPA AND THE MOST LIKELY ROOT CAUSE HAS BEEN IDENTIFIED. AS A RESULT, CORRECTIVE ACTIONS AND PROCEDURES ARE BEING IMPLEMENTED TO MITIGATE FURTHER OCCURRENCES. ROOT CAUSE DESCRIPTION: A CAPA WAS CONDUCTED TO DOCUMENT FURTHER INVESTIGATION AND ROOT CAUSE ANALYSIS RELATING TO THIS ISSUE. THE INVESTIGATION HAS IDENTIFED THE MOST LIKELY ROOT CAUSES AND CORRECTIVE ACTIONS ARE IN THE PROCESS OF BEING IMPLEMENTED. RATIONALE: BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE. THE INVESTIGATION HAS IDENTIFIED POTENTIAL ROOT CAUSE(S) FOR THIS ISSUE AND CORRECTIVE ACTIONS ARE IN THE PROCESS OF BEING IMPLEMENTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES GEL DID NOT MIGRATE PROPERLY. 6 TUBES DID NOT MIGRATE PROPERLY. NO REPORT OF MEDICAL INTERVENTIONS OR SERIOUS INJURY.

Additional Manufacturer Narrative · 1

MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8236728, MEDICAL DEVICE EXPIRATION DATE: 2019-08-31, DEVICE MANUFACTURE DATE: 2018-08-24. MEDICAL DEVICE LOT #: 8222847, MEDICAL DEVICE EXPIRATION DATE: 2019-07-31, DEVICE MANUFACTURE DATE: 2018-08-10. MEDICAL DEVICE LOT #: 7313631, MEDICAL DEVICE EXPIRATION DATE: 2018-10-31, DEVICE MANUFACTURE DATE: 2017-11-09. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES GEL DID NOT MIGRATE PROPERLY. 6 TUBES DID NOT MIGRATE PROPERLY. NO REPORT OF MEDICAL INTERVENTIONS OR SERIOUS INJURY.