FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 8065354 · Received November 13, 2018

Report

Report Number
1645337-2018-06799
Event Type
Injury
Date Received
November 13, 2018
Date of Event
October 17, 2018
Report Date
October 23, 2018
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001287
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION SUMMARY: UPON RECEIPT BY MENTOR, THE DEVICE CONTAINED NO FLUID. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE. WHITE MATERIAL WAS OBSERVED ON THE SHELL SURFACE. DURING VISUAL EXAMINATION OF THE DEVICE, MENTOR PRODUCT ANALYSIS LAB OBSERVED A CREASE EXTENDING TO BOTH VIEWS. A RENT WAS DISCOVERED WITHIN THE CREASE ON THE POSTERIOR VIEW THAT EXTENDED TO THE ANTERIOR VIEW, MEASURING APPROXIMATELY 0.7 CM. NO OTHER ANOMALIES WERE FOUND. MENTOR PRODUCTS ARE 100% VISUALLY INSPECTED PRIOR TO RELEASE IN ADDITION TO THOROUGH IN-PROCESS TESTING DURING SEVERAL STAGES OF THE MANUFACTURING PROCESS. MENTOR PRODUCT ANALYSIS LAB CONCLUDED THAT THE RENT AND CREASE OCCURRED SOMETIME SUBSEQUENT TO THE REMOVAL OF THE DEVICE FROM ITS PROTECTIVE PACKAGING. THE COMPLAINT OF DEFLATION WAS CONFIRMED SINCE A RENT WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE OR THE ORIGIN OF THE YELLOW AND WHITE MATERIAL OBSERVED. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. THESE FINDINGS ARE CONSISTENT WITH A CREASE/FOLD FAILURE WHICH IS A KNOWN INHERENT RISK ASSOCIATED WITH THE USE OF SALINE-FILLED MAMMARY PROSTHESES AND MAY BE THE RESULT OF ONE OR MORE OF THE FOLLOWING CONTRIBUTING FACTORS: UNDERINFLATION OR OVERINFLATION OF THE DEVICE, CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE - SUCH AS CAPSULAR CONTRACTURE OR TOO SMALL A BREAST POCKET, AND FOLDING OR WRINKLING OF THE SHELL IN THE BREAST POCKET. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON (B)(6) 2018, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A DEVICE HISTORY RECORD REVIEW IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ADDITIONAL CONCOMITANT MEDICAL PRODUCTS: SALINE MENTOR SMOOTH ROUND MODERATE PROFILE 425CC, CATALOG NUMBER 3501670, LOT NUMBER 204710. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION PRIMARY WITH A SALINE MENTOR SMOOTH ROUND MODERATE PROFILE 425CC BREAST IMPLANT AND EXPERIENCED DEFLATION ON THE RIGHT SIDE. DEFLATION WAS CONFIRMED BY PHYSICIAN DURING EXAM. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REMOVAL AND REPLACEMENT WITH SALINE MENTOR SMOOTH ROUND MODERATE PLUS PROFILE 450CC, CATALOG NUMBER 3502450, (B)(4), LOT NUMBER 7502179 (R) AND (B)(4), LOT NUMBER 6982509 (L) ON (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
903567 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 212142 00081317001287

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention