MENTOR SILTEX ROUND MODERATE PROFILE
Report
- Report Number
- 1645337-2018-06779
- Event Type
- Injury
- Date Received
- November 12, 2018
- Date of Event
- October 12, 2018
- Report Date
- October 15, 2018
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- UDI-DI
- 00081317001805
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE HISTORY RECORD (DHR) OF LOT NUMBER 232548 WAS REVIEWED ON (B)(6) 2018 AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THE EVALUATION OF THE RETURNED DEVICE WAS COMPLETED BY THE FAILURE ANALYSIS LAB ON (B)(6) 2018. IT WAS REPORTED THAT A (B)(6) FEMALE UNDERWENT PRIMARY BREAST AUGMENTATION WITH A MENTOR SILTEX ROUND MODERATE PROFILE 350CC SALINE PROSTHESIS. DEFLATION OF AN UNKNOWN SIDE WAS REPORTED POST PROCEDURE. UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE AND ON THE SHELL SURFACE. THE MENTOR PRODUCT ANALYSIS LAB DISCOVERED A RENT MEASURING APPROXIMATELY 0.6 CM ON THE POSTERIOR ASPECT. MICROSCOPIC EXAMINATION OF THE EDGES OF THE RENT GAVE NO INDICATIONS AS TO CAUSE. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE OR THE ETIOLOGY OF THE YELLOW MATERIAL FOUND ON THE DEVICE. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BECAUSE MENTOR PRODUCT ANALYSIS LAB WAS UNABLE TO CONFIRM ANY UNUSUAL FAILURE MODE, NO CORRECTIVE ACTION WILL BE TAKEN AT THIS TIME. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. (B)(4).
THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. A DEVICE HISTORY RECORD REVIEW IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: MENTOR SILTEX ROUND MODERATE PROFILE 350CC SALINE PROSTHESIS, CATALOG #3542655, LOT #223718. (B)(4).
IT WAS REPORTED THAT A (B)(6) -YEAR-OLD FEMALE UNDERWENT PRIMARY BREAST AUGMENTATION WITH A MENTOR SILTEX ROUND MODERATE PROFILE 350CC SALINE PROSTHESIS. DEFLATION OF AN UNKNOWN SIDE WAS REPORTED POST PROCEDURE. AS A RESULT, THE DEVICE WAS EXPLANTED ON (B)(6) 2018.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 901067 | MENTOR SILTEX ROUND MODERATE PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 00081317001805 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR | Required Intervention |