FDA Adverse Event Injury Summary report: N

MENTOR SILTEX ROUND MODERATE PROFILE

MDR report key: 8061671 · Received November 12, 2018

Report

Report Number
1645337-2018-06779
Event Type
Injury
Date Received
November 12, 2018
Date of Event
October 12, 2018
Report Date
October 15, 2018
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001805
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HISTORY RECORD (DHR) OF LOT NUMBER 232548 WAS REVIEWED ON (B)(6) 2018 AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THE EVALUATION OF THE RETURNED DEVICE WAS COMPLETED BY THE FAILURE ANALYSIS LAB ON (B)(6) 2018. IT WAS REPORTED THAT A (B)(6) FEMALE UNDERWENT PRIMARY BREAST AUGMENTATION WITH A MENTOR SILTEX ROUND MODERATE PROFILE 350CC SALINE PROSTHESIS. DEFLATION OF AN UNKNOWN SIDE WAS REPORTED POST PROCEDURE. UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE AND ON THE SHELL SURFACE. THE MENTOR PRODUCT ANALYSIS LAB DISCOVERED A RENT MEASURING APPROXIMATELY 0.6 CM ON THE POSTERIOR ASPECT. MICROSCOPIC EXAMINATION OF THE EDGES OF THE RENT GAVE NO INDICATIONS AS TO CAUSE. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE OR THE ETIOLOGY OF THE YELLOW MATERIAL FOUND ON THE DEVICE. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BECAUSE MENTOR PRODUCT ANALYSIS LAB WAS UNABLE TO CONFIRM ANY UNUSUAL FAILURE MODE, NO CORRECTIVE ACTION WILL BE TAKEN AT THIS TIME. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. (B)(4).

Additional Manufacturer Narrative · 1

THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. A DEVICE HISTORY RECORD REVIEW IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: MENTOR SILTEX ROUND MODERATE PROFILE 350CC SALINE PROSTHESIS, CATALOG #3542655, LOT #223718. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) -YEAR-OLD FEMALE UNDERWENT PRIMARY BREAST AUGMENTATION WITH A MENTOR SILTEX ROUND MODERATE PROFILE 350CC SALINE PROSTHESIS. DEFLATION OF AN UNKNOWN SIDE WAS REPORTED POST PROCEDURE. AS A RESULT, THE DEVICE WAS EXPLANTED ON (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
901067 MENTOR SILTEX ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 00081317001805

Patients

Seq Age Sex Outcome Treatment
1 39 YR Required Intervention